| Literature DB >> 15348922 |
R Schäfer1, U Soltész, P F Bernard.
Abstract
The head of an implanted hip joint endoprosthesis is exposed to torques, which are transferred during gait due to the friction between the head and the cup prosthesis. In prostheses with ceramic ball heads, which are widely used now, and in which the head is fixed onto the stem by conical clamping, these torques could possibly affect the connection. In this study, torques transferred from the cup to the head are compared to the torques which are required to loosen the head from the metallic spigot. The results show that for the investigated head and taper types and sizes, under normal conditions the connection is safe with respect to undesired rotation. However, it is shown that for polluted sliding surfaces the fixation strength could possibly be exceeded. Copyright 1998 Kluwer Academic PublishersEntities:
Year: 1998 PMID: 15348922 DOI: 10.1023/a:1008926213433
Source DB: PubMed Journal: J Mater Sci Mater Med ISSN: 0957-4530 Impact factor: 3.896