Literature DB >> 15347564

Automatic control of the inspired oxygen fraction in preterm infants: a randomized crossover trial.

Michael S Urschitz1, Werner Horn, Andreas Seyfang, Antonella Hallenberger, Tina Herberts, Silvia Miksch, Christian Popow, Ingo Müller-Hansen, Christian F Poets.   

Abstract

In preterm infants receiving supplemental oxygen, manual control of the inspired oxygen fraction is often time-consuming and inappropriate. We developed a system for automatic oxygen control and hypothesized that this system is more effective than routine manual oxygen control in maintaining target arterial oxygen saturation levels. We performed a randomized controlled crossover clinical trial in 12 preterm infants receiving nasal continuous positive airway pressure and supplemental oxygen. Periods with automatic and routine manual oxygen control were compared with periods of optimal control by a fully dedicated person. The median (range) percentage of time with arterial oxygen saturation levels within target range (87-96%) was 81.7% (39.0-99.8) for routine manual oxygen control, 91.0% (41.4-99.3) for optimal control, and 90.5% (59.0-99.4) for automatic control (ANOVA: p = 0.01). Pairwise post hoc comparisons revealed a statistically significant difference between automatic and routine manual oxygen control (Dunnett's test: p = 0.02). The frequency of manual oxygen adjustments was lowest in automatic control (Friedman's test: p < 0.001). Automatic oxygen control may optimize oxygen administration to preterm infants receiving nasal continuous positive airway pressure and reduce nursing time spent with oxygen control.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15347564     DOI: 10.1164/rccm.200407-929OC

Source DB:  PubMed          Journal:  Am J Respir Crit Care Med        ISSN: 1073-449X            Impact factor:   21.405


  22 in total

Review 1.  The design of future pediatric mechanical ventilation trials for acute lung injury.

Authors:  Robinder G Khemani; Christopher J L Newth
Journal:  Am J Respir Crit Care Med       Date:  2010-08-23       Impact factor: 21.405

2.  Closed loop mechanical ventilation.

Authors:  Marc Wysocki; Philippe Jouvet; Samir Jaber
Journal:  J Clin Monit Comput       Date:  2013-04-07       Impact factor: 2.502

3.  Hypoxic events and concomitant factors in preterm infants on non-invasive ventilation.

Authors:  Omid Sadeghi Fathabadi; Timothy Gale; Kevin Wheeler; Gemma Plottier; Louise S Owen; J C Olivier; Peter A Dargaville
Journal:  J Clin Monit Comput       Date:  2016-02-20       Impact factor: 2.502

4.  A higher incidence of intermittent hypoxemic episodes is associated with severe retinopathy of prematurity.

Authors:  Juliann M Di Fiore; Jeffrey N Bloom; Faruk Orge; Alison Schutt; Mark Schluchter; Vinay K Cheruvu; Michele Walsh; Neil Finer; Richard J Martin
Journal:  J Pediatr       Date:  2010-03-20       Impact factor: 4.406

5.  Clinical Evaluation of an Automatic Oxygen Control System for Premature Infants Receiving High-Flow Nasal Cannula for Respiratory Support: A Pilot Study.

Authors:  Xuefeng Hou; Akram Faqeeh; Ramak Amjad; John Pardalos; Roger Fales
Journal:  J Med Device       Date:  2022-05-10       Impact factor: 0.743

Review 6.  Automated Oxygen Delivery in Neonatal Intensive Care.

Authors:  Vrinda Nair; Prakash Loganathan; Mithilesh Kumar Lal; Thomas Bachman
Journal:  Front Pediatr       Date:  2022-06-22       Impact factor: 3.569

7.  Fully automated predictive intelligent control of oxygenation (PRICO) in resuscitation and ventilation of preterm lambs.

Authors:  Matthias C Hütten; Tom G Goos; Daan Ophelders; Maria Nikiforou; Elke Kuypers; Monique Willems; Hendrik J Niemarkt; Jenny Dankelman; Peter Andriessen; Thilo Mohns; Irwin K M Reiss; Boris W Kramer
Journal:  Pediatr Res       Date:  2015-08-31       Impact factor: 3.756

8.  Maintaining optimal oxygen saturation in premature infants.

Authors:  Yoke Yen Lau; Yih Yann Tay; Varsha Atul Shah; Pisun Chang; Khuan Tai Loh
Journal:  Perm J       Date:  2011

9.  A pilot prospective study on closed loop controlled ventilation and oxygenation in ventilated children during the weaning phase.

Authors:  Philippe Jouvet; Allen Eddington; Valérie Payen; Alice Bordessoule; Guillaume Emeriaud; Ricardo Lopez Gasco; Marc Wysocki
Journal:  Crit Care       Date:  2012-05-16       Impact factor: 9.097

10.  Feasibility and reliability of an automated controller of inspired oxygen concentration during mechanical ventilation.

Authors:  Kaouther Saihi; Jean-Christophe M Richard; Xavier Gonin; Thomas Krüger; Michel Dojat; Laurent Brochard
Journal:  Crit Care       Date:  2014-02-19       Impact factor: 9.097

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.