Side-to-side sutureless vascular anastomosis with magnets.

Detlev Erdmann1, Ranya Sweis, Christoph Heitmann, Koji Yasui, Kevin C Olbrich, L Scott Levin, A Adam Sharkawy, Bruce Klitzman.   

Abstract

OBJECTIVE: Abbe and Payr introduced vascular techniques and devices to facilitate vessel anastomosis over a century ago. Obora published the idea of a sutureless vascular anastomosis with use of magnetic rings in 1978. The purpose of this study was to assess the performance of a new magnetic device to perform a side-to-side arteriovenous anastomosis in a dog model.
MATERIAL AND METHODS: Male fox hounds (25 kg) were treated preoperatively and daily postoperatively with clopidogrel bisulfate (Plavix) and aspirin. The femoral artery and vein were exposed unilaterally in 3 dogs and bilaterally in 4 dogs (n = 11 anastomoses). A 4-mm arteriotomy was performed, and 1 oval magnet 0.5 mm thick was inserted into the lumen of the artery and a second magnet was applied external to the artery, compressing and stabilizing the arterial wall to create a magnetic port. An identical venous magnetic port was created with another pair of oval magnets. When the 2 ports were allowed to approach each other, they self-aligned and magnetically coupled to complete the arteriovenous anastomosis. Patency was assessed for the first hour with direct observation, again after 9 weeks with duplex ultrasound scanning, and at 10 weeks under direct open observation. The anastomoses were explanted after 10 weeks. Hydrodynamic resistance was measured ex vivo on the final 8 anastomoses by measuring the pressure drop across an anastomosis with a known flow rate.
RESULTS: After implantation, very high flow created visible turbulence and palpable vibration. All 11 anastomoses were patent under direct observation and palpation. Ten of 11 anastomoses were clearly patent on duplex scans, and patency of 1 anastomosis was questionable. Hydrodynamic resistance averaged 0.73 +/- 0.33 mm Hg min/mL (mean +/- SEM).
CONCLUSIONS: Vascular anastomoses performed with magnets demonstrated feasibility; exhibited 100% patency after 10 weeks in a dog arteriovenous shunt model; lacked apparent aneurysm or other potentially catastrophic failure; demonstrated remodeling of the vessel wall after several weeks to incorporate the magnets, making the magnetic force unnecessary; and warrants further study in vessels with different sizes, flow rates, and locations. CLINICAL RELEVANCE: We present a magnet-based device used to perform side-to-side peripheral vascular anastomoses. Its advantages include the ability to anastomose vessels without requiring circumferential surgical exposure. Vascular anastomosis performed with these magnets demonstrated 100% patency in the dog, lacked apparent aneurysm or other potentially catastrophic failure, and demonstrated remodeling of the vessel wall after several weeks, to incorporate the magnets, making indefinite retention of field strength unnecessary. This technique could enable minimally invasive procedures, such as complex reconstructive and revascularizing surgery, and warrants further study in vessels with different sizes, flow rates, and locations.

Entities:  

Mesh:

Year:  2004        PMID: 15337881     DOI: 10.1016/j.jvs.2004.05.026

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  16 in total

Review 1.  Compression anastomoses in colorectal surgery: a review.

Authors:  A P Zbar; Y Nir; A Weizman; M Rabau; A Senagore
Journal:  Tech Coloproctol       Date:  2012-04-26       Impact factor: 3.781

2.  Magnetic compression gastrostomy in the rat.

Authors:  Ibrahim Uygun; Mehmet Hanifi Okur; Hasan Cimen; Aysenur Keles; Ozben Yalcin; Hayrettin Ozturk; Selcuk Otcu
Journal:  Pediatr Surg Int       Date:  2012-01-24       Impact factor: 1.827

3.  Retraction by surface ferromagnetisation of target tissues: preliminary studies on feasibility of magnetic retraction for endoscopic surgery.

Authors:  Zhigang Wang; Lijun Wang; Benjie Tang; Timothy Frank; Stuart Brown; Alfred Cuschieri
Journal:  Surg Endosc       Date:  2007-12-11       Impact factor: 4.584

4.  Magnamosis: a novel technique for the management of rectal atresia.

Authors:  Katie W Russell; Michael D Rollins; G Peter Feola; Eric R Scaife
Journal:  BMJ Case Rep       Date:  2014-08-05

5.  Innovative magnetic rings for circumferential mucosectomy: preliminary research.

Authors:  Geng Shu An; Zhang Qing Huai; Zhou Sheng; Zhang Shu Min; Yang Jing Wen; Hao Dong Ming; Zhang Zuo Xing
Journal:  Surg Today       Date:  2015-01       Impact factor: 2.549

6.  Magnetic Anastomosis Rings to Create Portacaval Shunt in a Canine Model of Portal Hypertension.

Authors:  Hao-Hua Wang; Jia Ma; Shan-Pei Wang; Feng Ma; Jian-Wen Lu; Xiang-Hua Xu; Yi Lv; Xiao-Peng Yan
Journal:  J Gastrointest Surg       Date:  2018-08-21       Impact factor: 3.452

7.  Treatment of rectovaginal fistula by magnetic compression.

Authors:  Zhan-Fei She; Xiao-Peng Yan; Feng Ma; Hao-Hua Wang; Huan Yang; Ai-Hua Shi; Liang Wang; Xin Qi; Bin Xiao; Yu-Liang Zou; Yi Lv
Journal:  Int Urogynecol J       Date:  2016-08-16       Impact factor: 2.894

8.  Vascular anastomosis using controlled phase transitions in poloxamer gels.

Authors:  Edward I Chang; Michael G Galvez; Jason P Glotzbach; Cynthia D Hamou; Samyra El-ftesi; C Travis Rappleye; Kristin-Maria Sommer; Jayakumar Rajadas; Oscar J Abilez; Gerald G Fuller; Michael T Longaker; Geoffrey C Gurtner
Journal:  Nat Med       Date:  2011-08-28       Impact factor: 53.440

9.  Creation of transcatheter aortopulmonary and cavopulmonary shunts using magnetic catheters: feasibility study in swine.

Authors:  Daniel S Levi; Saar Danon; Brent Gordon; Nicky Virdone; Fernando Vinuela; Sanjay Shah; Greg Carman; John W Moore
Journal:  Pediatr Cardiol       Date:  2009-04-14       Impact factor: 1.655

10.  Ingestion of magnetic foreign bodies causing multiple bowel perforations.

Authors:  Moritz F Kircher; Sarah Milla; Michael J Callahan
Journal:  Pediatr Radiol       Date:  2007-07-11
View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.