| Literature DB >> 15330032 |
G Pizà1, V L Caja, M A González-Viejo, A Navarro.
Abstract
Pin loosening and infection are inherent complications of external fixation. This study deals with their effects of using either hydroxyapatite (HA)-coated or uncoated external fixation pins in leg-lengthening procedures on patients of short stature. We used HA-coated pins on one side and uncoated pins on the other (randomly determined) in 28 bilateral lengthenings undertaken in 23 patients. A total of 322 pins was used. The mean implantation time was 530 days and the mean lengthening achieved was 78% of initial bone length. Mean extraction torque was 7611.6 Nmm degree(-1) for HA-coated and 85.4 Nmm degree(-1) for uncoated pins (p < 0.001). The rate of pin loosening was 4% (7/161) for HA-coated and 80% (129/161) for uncoated pins (p < 0.001). There was no statistically significant difference in the incidence of pin-track infection between the two groups. The use of HA coating appears to be an effective method of reducing the incidence of pin loosening in external fixation with a long implantation time and for mechanically highly stressed procedures such as leg lengthening for short stature.Entities:
Mesh:
Substances:
Year: 2004 PMID: 15330032 DOI: 10.1302/0301-620x.86b6.13875
Source DB: PubMed Journal: J Bone Joint Surg Br ISSN: 0301-620X