Evaluation of accuracy and reproducibility of peak flowmeters at 1,400 m.

R M Gardner1, R O Crapo, B R Jackson, R L Jensen.   

Abstract

Peak flow meters provide physicians and patients with objective measures about changes in pulmonary obstruction. We evaluated eight models of peak flowmeters and measured their accuracy and reproducibility with methods recently recommended by the National Asthma Education Program (NAEP). Waveforms from the American Thoracic Society's spirometer testing set were used to drive a computer-controlled syringe. Testing was done at Salt Lake City at an altitude 1,400 m. It appears that the original Wright peak flowmeter has been used as the "de facto" standard. We found that the original Wright peak flowmeter overestimated flows in its midrange; and, as a consequence, most of the other peak flowmeters also overestimated peak flows. The overestimation of peak flows may have been understated because of the 1,400-m altitude testing site. To the credit of the instrument manufacturers, we were pleasantly surprised with the quality, accuracy, and reproducibility of presently available peak flowmeters; however, as a result of our testing, we suspect that with little effort, manufacturers of peak flowmeters could improve the accuracy of their devices. Standardized testing methods and equipment should make the task of peak flowmeter design, manufacture, and testing even easier. We trust that manufacturers of peak flowmeters will respond appropriately and improve their instruments.

Entities:  

Mesh:

Year:  1992        PMID: 1532549     DOI: 10.1378/chest.101.4.948

Source DB:  PubMed          Journal:  Chest        ISSN: 0012-3692            Impact factor:   9.410


  6 in total

1.  Blow what's the flow?

Authors:  M U Sanklecha; K Raghavan
Journal:  Indian J Pediatr       Date:  1994 Jan-Feb       Impact factor: 1.967

2.  Peak flow meters: a problem of scale.

Authors:  M R Miller; P H Ouanjer
Journal:  BMJ       Date:  1994-02-26

3.  National Athletic Trainers' Association position statement: management of asthma in athletes.

Authors:  Michael G Miller; John M Weiler; Robert Baker; James Collins; Gilbert D'Alonzo
Journal:  J Athl Train       Date:  2005 Jul-Sep       Impact factor: 2.860

Review 4.  Brittle asthma.

Authors:  J G Ayres; J F Miles; P J Barnes
Journal:  Thorax       Date:  1998-04       Impact factor: 9.139

5.  Care of asthma. A personal approach.

Authors:  M M Dean
Journal:  Can Fam Physician       Date:  1993-11       Impact factor: 3.275

6.  Potential effects of correction of inaccuracies of the mini-Wright peak expiratory flow meter on the use of an asthma self-management plan.

Authors:  J F Miles; W Tunnicliffe; R M Cayton; J G Ayres; M R Miller
Journal:  Thorax       Date:  1996-04       Impact factor: 9.139

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.