An efficient calibration method for freehand 3-D ultrasound imaging systems.

Daniel F Leotta1.   

Abstract

A phantom has been developed to quickly calibrate a freehand 3-D ultrasound (US) imaging system. Calibration defines the spatial relationship between the US image plane and an external tracking device attached to the scanhead. The phantom consists of a planar array of strings and beads, and a set of out-of-plane strings that guide the user to proper scanhead orientation for imaging. When an US image plane is coincident with the plane defined by the strings, the calibration parameters are calculated by matching of homologous points in the image and phantom. The resulting precision and accuracy of the 3-D imaging system are similar to those achieved with a more complex calibration procedure. The 3-D reconstruction performance of the calibrated system is demonstrated with a magnetic tracking system, but the method could be applied to other tracking devices.

Mesh:

Year:  2004        PMID: 15313332     DOI: 10.1016/j.ultrasmedbio.2004.05.007

Source DB:  PubMed          Journal:  Ultrasound Med Biol        ISSN: 0301-5629            Impact factor:   2.998


  7 in total

Review 1.  New technologies for information retrieval to achieve situational awareness and higher patient safety in the surgical operating room: the MRI institutional approach and review of the literature.

Authors:  Michael Kranzfelder; Armin Schneider; Sonja Gillen; Hubertus Feussner
Journal:  Surg Endosc       Date:  2010-08-19       Impact factor: 4.584

2.  GPU based real-time instrument tracking with three-dimensional ultrasound.

Authors:  Paul M Novotny; Jeff A Stoll; Nikolay V Vasilyev; Pedro J del Nido; Pierre E Dupont; Todd E Zickler; Robert D Howe
Journal:  Med Image Anal       Date:  2007-07-05       Impact factor: 8.545

3.  Three-dimensional ultrasound scanning.

Authors:  Aaron Fenster; Grace Parraga; Jeff Bax
Journal:  Interface Focus       Date:  2011-06-01       Impact factor: 3.906

4.  Passive markers for tracking surgical instruments in real-time 3-D ultrasound imaging.

Authors:  Jeffrey Stoll; Hongliang Ren; Pierre E Dupont
Journal:  IEEE Trans Med Imaging       Date:  2011-10-25       Impact factor: 10.048

Review 5.  Robotics and imaging in congenital heart surgery.

Authors:  Nikolay V Vasilyev; Pierre E Dupont; Pedro J del Nido
Journal:  Future Cardiol       Date:  2012-03

6.  Verification of fetal brain responses by coregistration of fetal ultrasound and fetal magnetoencephalography data.

Authors:  C Micheli; J McCubbin; P Murphy; H Eswaran; C L Lowery; E Ortiz; H Preissl
Journal:  Neuroimage       Date:  2009-09-22       Impact factor: 6.556

7.  First Steps Toward Ultrasound-Based Motion Compensation for Imaging and Therapy: Calibration with an Optical System and 4D PET Imaging.

Authors:  Julia Schwaab; Christopher Kurz; Cristina Sarti; André Bongers; Frédéric Schoenahl; Christoph Bert; Jürgen Debus; Katia Parodi; Jürgen Walter Jenne
Journal:  Front Oncol       Date:  2015-11-30       Impact factor: 6.244

  7 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.