Literature DB >> 15302782

Beating heart catheter-based edge-to-edge mitral valve procedure in a porcine model: efficacy and healing response.

James I Fann1, Frederick G St Goar, Jan Komtebedde, Mehmet C Oz, Peter C Block, Elyse Foster, Jagdish Butany, Ted Feldman, Thomas A Burdon.   

Abstract

BACKGROUND: Surgical edge-to-edge repair has been used in the treatment of mitral regurgitation. We evaluated the ability of a catheter-delivered clip (Evalve, Inc) to achieve edge-to-edge mitral valve approximation without cardiopulmonary bypass and the healing response of this technique. METHODS AND
RESULTS: Twenty-one pigs underwent general anesthesia and left thoracotomy. A 10F flexible delivery catheter with a clip was placed into the left atrium. With echocardiographic and fluoroscopic guidance, the clip grasped and approximated the mid portion of the anterior and posterior leaflets. After a double orifice had been confirmed, the clip was detached and the catheter withdrawn. All animals survived and had successful clip placement. Three animals were euthanized at 4 weeks, 9 at 12 weeks, 1 at 17 weeks, 7 at 24 weeks, and 1 at 52 weeks. The clip was well positioned, with leaflet approximation in all animals except 1, in which the clip separated from the posterior leaflet at 4 weeks without affecting valve function. The clip was modified and implanted in 4 pigs; all were intact at 12 to 24 weeks. Scanning electron microscopy showed clip encapsulation with complete endothelialization. Mitral stenosis and thromboembolism did not develop. Two animals developed endocarditis (1 at 12 weeks and 1 at 17 weeks). Progressive healing occurred in all other animals.
CONCLUSIONS: Edge-to-edge mitral valve approximation can be successfully and reliably achieved with a catheter-delivered clip without cardiopulmonary bypass, resulting in durable healing. The success of this device supports the development of a percutaneous catheter-based system for mitral valve repair.

Entities:  

Mesh:

Year:  2004        PMID: 15302782     DOI: 10.1161/01.CIR.0000139855.12616.15

Source DB:  PubMed          Journal:  Circulation        ISSN: 0009-7322            Impact factor:   29.690


  20 in total

1.  Mitral valve analysis using a novel 3D holographic display: a feasibility study of 3D ultrasound data converted to a holographic screen.

Authors:  Jan Otto Beitnes; Lars Gunnar Klæboe; Jørn Skaarud Karlsen; Stig Urheim
Journal:  Int J Cardiovasc Imaging       Date:  2014-11-13       Impact factor: 2.357

Review 2.  Minimally invasive heart valve surgery: how and why in 2012.

Authors:  Rakesh M Suri; Nassir M Thalji
Journal:  Curr Cardiol Rep       Date:  2012-04       Impact factor: 2.931

3.  Current status of percutaneous valvular procedures.

Authors:  Dominique Himbert; Eric Brochet; David Messika-Zeitoun; Alec Vahanian
Journal:  Curr Treat Options Cardiovasc Med       Date:  2006-11

4.  Need for new materials, biofunctionalization and non-surgical heart valve technology.

Authors:  Jan Sochman
Journal:  World J Cardiol       Date:  2010-03-26

Review 5.  Using imaging to guide patient selection and performance of catheter-based mitral valve repair for mitral regurgitation.

Authors:  D Scott Lim
Journal:  J Cardiovasc Transl Res       Date:  2013-09-21       Impact factor: 4.132

6.  Percutaneous mitral valve repair using the edge-to-edge technique in a high-risk population.

Authors:  B J L Van den Branden; M C Post; M J Swaans; B J W M Rensing; F D Eefting; H W M Plokker; W Jaarsma; J A S Van der Heyden
Journal:  Neth Heart J       Date:  2010-09       Impact factor: 2.380

7.  Mitral cerclage annuloplasty, a novel transcatheter treatment for secondary mitral valve regurgitation: initial results in swine.

Authors:  June-Hong Kim; Ozgur Kocaturk; Cengizhan Ozturk; Anthony Z Faranesh; Merdim Sonmez; Smita Sampath; Christina E Saikus; Ann H Kim; Venkatesh K Raman; J Andrew Derbyshire; William H Schenke; Victor J Wright; Colin Berry; Elliot R McVeigh; Robert J Lederman
Journal:  J Am Coll Cardiol       Date:  2009-08-11       Impact factor: 24.094

8.  Percutaneous Mitral Valve Edge-to-Edge Repair for Degenerative Mitral Regurgitation.

Authors:  Jacob P Dal-Bianco; Ignacio Inglessis; Serguei Melnitchouk; Maureen Daher; Igor F Palacios
Journal:  Curr Treat Options Cardiovasc Med       Date:  2015-07

Review 9.  The Evolving Role of Percutaneous Mitral Valve Repair.

Authors:  Merrill H Stewart; J Stephen Jenkins
Journal:  Ochsner J       Date:  2016

10.  Emerging approaches of transcatheter valve repair/insertion.

Authors:  Maurizio Taramasso; Micaela Cioni; Andrea Giacomini; Iassen Michev; Cosmo Godino; Matteo Montorfano; Antonio Colombo; Ottavio Alfieri; Francesco Maisano
Journal:  Cardiol Res Pract       Date:  2010-07-25       Impact factor: 1.866

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.