Literature DB >> 1529874

Balloon catheter with check valves for experimental relief of acute aortic regurgitation.

T Matsubara1, M Yamazoe, Y Tamura, M Ohshima, Y Yamazaki, M Suzuki, T Izumi, A Shibata.   

Abstract

To evaluate the ability of a newly designed balloon catheter with check valves to temporarily relieve hemodynamic deterioration in acute aortic regurgitation, we produced an experimental model of acute aortic regurgitation in closed-chest dogs using endomyocardial biopsy forceps. Aortic regurgitation was produced until an increase in aortic pulse pressure of over 50% was achieved. Left ventricular end-diastolic pressure rapidly increased after the production of aortic regurgitation. Immediately after the catheter began functioning, pulse pressure decreased from 133 +/- 1 to 78 +/- 5 mm Hg (mean +/- SEM) and left ventricular end-diastolic pressure also decreased from 26 +/- 1 to 13 +/- 1 mm Hg. These effects lasted as long as the catheter functioned. Although a mild (21 +/- 8 mm Hg) pressure gradient between the left ventricular peak systolic pressure and the aortic peak systolic pressure was observed when this catheter was used, forward stroke volume was no less than in the group in which the catheter had not been used. These findings suggest that the balloon catheter with check valves may effectively reduce aortic regurgitation.

Entities:  

Mesh:

Year:  1992        PMID: 1529874     DOI: 10.1016/0002-8703(92)90984-4

Source DB:  PubMed          Journal:  Am Heart J        ISSN: 0002-8703            Impact factor:   4.749


  5 in total

Review 1.  Which strategy for a protein crystallization project?

Authors:  C E Kundrot
Journal:  Cell Mol Life Sci       Date:  2004-03       Impact factor: 9.261

2.  First human case of retrograde transcatheter implantation of an aortic valve prosthesis.

Authors:  David Paniagua; José A Condado; José Besso; Manuel Vélez; Bruno Burger; Salvatore Bibbo; Douglas Cedeno; Harry Acquatella; Carlos Mejia; Eduardo Induni; R David Fish
Journal:  Tex Heart Inst J       Date:  2005

Review 3.  History of percutaneous aortic valve prosthesis.

Authors:  Henning Rud Andersen
Journal:  Herz       Date:  2009-08       Impact factor: 1.443

Review 4.  [Transcatheter aortic valve replacement: when should it be used in aortic stenosis?].

Authors:  Lukas Jaroslaw Motloch; Sara Reda; Dennis Rottlaender; Matthias Heigert; Uta C Hoppe
Journal:  Wien Med Wochenschr       Date:  2012-08-03

Review 5.  [Transcatheter-based aortic valve implantation. Present and future technologies].

Authors:  Stefan Sack; Joachim Schofer
Journal:  Herz       Date:  2009-08       Impact factor: 1.443

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.