Ultrasound navigation through the heart for off-pump aortic valved stent implantation: new tools for new goals.

Christoph H Huber1, Mohammed Nasratulla, Monique Augstburger, Ludwig K von Segesser.   

Abstract

PURPOSE: To validate the use of simultaneous intracardiac and intravascular ultrasound (IVUS) guidance for periprocedural dimension assessment, Residual-Coronary/Sinus-Stent Index (RCSSI) estimation, target site identification, deployment monitoring, and quality control of off-pump aortic Valved Stent implantation.
METHODS: Five pigs (56+/-5 kg) underwent off-pump orthotopic aortic valve implantation using a custom-made self-expanding Valved Stent. Intracardiac ultrasound (AcuNav) was introduced via the right femoral vein. After left-sided thoracotomy, pursestring sutures were placed on the left ventricular apex. Following heparinization, a guidewire was inserted through the apex and advanced over the aortic valve under fluoroscopy. A wire-guided IVUS catheter transducer (6-F, 12.5-MHz) was inserted and the aortic target site identified. IVUS probe location was tracked with AcuNav, and measures of the aortic root were taken by both. After removal of the IVUS, the Valved Stent delivery system was introduced over the guidewire under fluoroscopy and AcuNav monitoring; the Valved Stent was deployed over the native valves. In vivo assessment included leaflet motion, planimetric valve orifice and RCSSI (stent to aortic wall distance/coronary diameter) calculations, coronary blood flow characteristics, transvalvular gradient, regurgitation, and paravalvular leaking in combination with continuous cardiac output measures. Macroscopic analysis was performed at necropsy.
RESULTS: IVUS dimensions of the aortic root were equal to AcuNav and necropsy dimensions. Both tools showed good valvular function, with full valvular opening and closing in 3 of 5 valves. At necropsy, the 3 aortic Valved Stents were safely anchored. One Valved Stent was placed supra-annularly; 2 dislodged into the left ventricle because of size mismatch. One Valved Stent showed a moderate to severe paravalvular leak.
CONCLUSIONS: Simultaneous intracardiac and intravascular ultrasound-guided off-pump orthotopic aortic Valved Stent implantation via left sided thoracotomy is feasible in an animal model. IVUS and intracardiac ultrasound allow adequate aortic dimension assessment and Valved Stent delivery monitoring, as well as postimplantation coronary flow evaluation.

Entities:  

Mesh:

Year:  2004        PMID: 15298500     DOI: 10.1583/04-1240.1

Source DB:  PubMed          Journal:  J Endovasc Ther        ISSN: 1526-6028            Impact factor:   3.487


  4 in total

1.  Challenges and implementation of radiation-force imaging with an intracardiac ultrasound transducer.

Authors:  Stephen J Hsu; Brian J Fahey; Douglas M Dumont; Patrick D Wolf; Gregg E Trahey
Journal:  IEEE Trans Ultrason Ferroelectr Freq Control       Date:  2007-05       Impact factor: 2.725

2.  Transapical access closure: the TA PLUG device.

Authors:  Henriette Brinks; Fabian Nietlispach; Volkhard Göber; Lars Englberger; Peter Wenaweser; Bernhard Meier; Thierry Carrel; Christoph Huber
Journal:  Interact Cardiovasc Thorac Surg       Date:  2013-07-09

3.  A pilot study on magnetic navigation for transcatheter aortic valve implantation using dynamic aortic model and US image guidance.

Authors:  Zhe Luo; Junfeng Cai; Lixu Gu
Journal:  Int J Comput Assist Radiol Surg       Date:  2013-01-11       Impact factor: 2.924

4.  Real-time aortic valve segmentation from transesophageal echocardiography sequence.

Authors:  Junfeng Cai; Xiahai Zhuang; Yuanyuan Nie; Zhe Luo; Lixu Gu
Journal:  Int J Comput Assist Radiol Surg       Date:  2014-08-03       Impact factor: 2.924

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.