A comparison of human muscle temperature increases during 3-MHz continuous and pulsed ultrasound with equivalent temporal average intensities.

Joseph A Gallo1, David O Draper, Lori Thein Brody, Gilbert W Fellingham.   

Abstract

STUDY
DESIGN: A repeated-measure crossover design was used. The independent variable was the type of ultrasound (pulsed or continuous) and the dependent variable was intramuscular temperature.
OBJECTIVE: To compare changes in intramuscular temperature resulting from the use of pulsed ultrasound versus continuous ultrasound with an equivalent spatial average temporal average (SATA) intensity.
BACKGROUND: There is a lack of research on the heat-generating capabilities of pulsed ultrasound within human muscle. METHODS AND MEASURES: The subjects were 16 healthy volunteers (mean age +/- SD, 21.3 +/- 2.5 years). Each subject was treated with pulsed ultrasound (3 MHz, 1.0 W/cm2, 50% duty cycle, for 10 minutes) and continuous ultrasound (3 MHz, 0.5 W/cm2, for 10 minutes) during a single testing session. Tissue temperature returned to baseline and stabilized between treatments and treatment order was randomized. Tissue temperature was measured every 30 seconds using a 26-gauge needle microprobe inserted at a depth of 2 cm in the left medial gastrocnemius muscle. Data were analyzed using a linear mixed model.
RESULTS: Treatment with continuous ultrasound produced a mean (+/-SD) temperature increase of 2.8 degrees C +/- 0.8 degrees C above baseline. Treatment with pulsed ultrasound produced a mean (+/-SD) temperature increase of 2.8 degrees C +/- 0.7 degrees C above baseline. Statistical analysis revealed no significant differences in either the extent or rate of temperature increases between the 2 modes of ultrasound application.
CONCLUSION: Pulsed ultrasound (3 MHz, 1.0 W/cm2, 50% duty cycle, for 10 minutes) produces similar intramuscular temperature increases as continuous ultrasound (3 MHz, 0.5 W/cm2, for 10 minutes) at a 2-cm depth in the human gastrocnemius. Spatial average temporal average intensity is an important consideration when selecting pulsed ultrasound parameters intended to deliver nonthermal effects.

Entities:  

Mesh:

Year:  2004        PMID: 15296367     DOI: 10.2519/jospt.2004.34.7.395

Source DB:  PubMed          Journal:  J Orthop Sports Phys Ther        ISSN: 0190-6011            Impact factor:   4.751


  11 in total

1.  Cell separation using tilted-angle standing surface acoustic waves.

Authors:  Xiaoyun Ding; Zhangli Peng; Sz-Chin Steven Lin; Michela Geri; Sixing Li; Peng Li; Yuchao Chen; Ming Dao; Subra Suresh; Tony Jun Huang
Journal:  Proc Natl Acad Sci U S A       Date:  2014-08-25       Impact factor: 11.205

2.  Dexamethasone Sodium Phosphate Penetration During Phonophoresis at 2 Ultrasound Frequencies.

Authors:  Justin H Rigby; Austin M Hagan; Austin R Kelcher; Chang Ji
Journal:  J Athl Train       Date:  2020-06-23       Impact factor: 2.860

3.  The uncertainty (validity and reliability) of three electrothermometers in therapeutic modality research.

Authors:  Lisa S Jutte; Kenneth L Knight; Blaine C Long; Jeremy R Hawkins; Shane S Schulthies; Ethan B Dalley
Journal:  J Athl Train       Date:  2005 Jul-Sep       Impact factor: 2.860

4.  Ultrasound-induced thermal elevation in clotted blood and cranial bone.

Authors:  Volodymyr Nahirnyak; T Douglas Mast; Christy K Holland
Journal:  Ultrasound Med Biol       Date:  2007-05-09       Impact factor: 2.998

5.  Intramuscular temperature differences between the mid-point and peripheral effective radiating area with ultrasound.

Authors:  Michale G Miller; Janae R Longoria; Christopher C Cheatham; Robert J Baker; Timothy J Michael
Journal:  J Sports Sci Med       Date:  2008-06-01       Impact factor: 2.988

6.  Blisters on the anterior shin in 3 research subjects after a 1-MHz, 1.5-W/cm , continuous ultrasound treatment: a case series.

Authors:  Jamie L Frye; Lennart D Johns; James A Tom; Christopher D Ingersoll
Journal:  J Athl Train       Date:  2007 Jul-Sep       Impact factor: 2.860

7.  Phonophoresis and the absorption of dexamethasone in the presence of an occlusive dressing.

Authors:  Susan Saliba; Dilaawar J Mistry; David H Perrin; Joe Gieck; Arthur Weltman
Journal:  J Athl Train       Date:  2007 Jul-Sep       Impact factor: 2.860

8.  Overview of ultrasound usage trends in orthopedic and sports physiotherapy.

Authors:  Wouber Herickson de Brito Vieira; Kardec Alecxandro Aguiar; Kimberly Moreira da Silva; Pablo Miranda Canela; Flávio Santos da Silva; Bento João Abreu
Journal:  Crit Ultrasound J       Date:  2012-05-28

9.  Temperature changes caused by the difference in the distance between the ultrasound transducer and bone during 1 mhz and 3 mhz continuous ultrasound: a phantom study.

Authors:  Akihiko Ohwatashi; Satoshi Ikeda; Katsuhiro Harada; Yurie Kamikawa; Akira Yoshida; Kazuhiro Inoue; Nobuhiko Yanagida; Kiyohiro Fukudome; Ryoji Kiyama; Tadasu Ohshige; Tetsuo Maeda
Journal:  J Phys Ther Sci       Date:  2015-01-09

10.  The effect of continuous ultrasound on chronic non-specific low back pain: a single blind placebo-controlled randomized trial.

Authors:  Safoora Ebadi; Noureddin Nakhostin Ansari; Soofia Naghdi; Shohre Jalaei; Mirmostafa Sadat; Hosein Bagheri; Maurits W Vantulder; Nicholas Henschke; Ehsan Fallah
Journal:  BMC Musculoskelet Disord       Date:  2012-10-02       Impact factor: 2.362

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.