Literature DB >> 15290201

[Non-contact donor cornea trephination with a flying spot excimer laser system].

M Müller1, Z Sherif, U Pleyer, C Hartmann.   

Abstract

BACKGROUND: One of the most common complications after mechanical penetrating keratoplasty is the occurrence of irregular astigmatism. We developed a method to prepare donor grafts for transplantation applying excimer laser techniques. The quality of the cut edges and exposure time were evaluated.
MATERIALS AND METHODS: The beam of an experimental excimer laser (OPTex, Lambda Physik) was homogenized and guided by an optical system including lenses and scanner mirrors. A special software was used to conduct the laser beam in a circular mode over the surface of the cornea. Corneas from porcine eyes were stabilized in an artificial anterior chamber. Twenty in vitro explants were trephinated applying our method and subsequently investigated by electron microscopy and histology.
RESULTS: All experiments provided regular and smooth cut edges. The average exposure time until perforation was 10.5 min (SD 2.7 min), the diameter of the grafts was 8.0 mm.
CONCLUSION: The experiments showed the possibility of non-contact trephination and generation of smooth cut edges with a rotating focussed excimer laser beam in a porcine cornea model. A major advantage of the system is the possibility of customized "tailored" grafts.

Entities:  

Mesh:

Year:  2005        PMID: 15290201     DOI: 10.1007/s00347-004-1079-0

Source DB:  PubMed          Journal:  Ophthalmologe        ISSN: 0941-293X            Impact factor:   1.059


  14 in total

1.  Graft endothelium and thickness after penetrating keratoplasty, comparing mechanical and excimer laser trephination: a prospective randomised study.

Authors:  B Seitz; A Langenbucher; N X Nguyen; M M Kus; M Küchle; G O Naumann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2001-01       Impact factor: 3.117

2.  Experimental 193-nm excimer laser trephination with divergent cut angles in penetrating keratoplasty.

Authors:  B Seitz; A Behrens; A Langenbucher; M M Kus; G O Naumann
Journal:  Cornea       Date:  1998-07       Impact factor: 2.651

3.  Effect of trephination technique on the ultrastructure of corneal transplants: guided trephine system v posterior punch technique.

Authors:  W Radner; C Skorpik; R Loewe; C Mudrich; G Radner; R Mallinger
Journal:  Br J Ophthalmol       Date:  1999-10       Impact factor: 4.638

4.  Nonmechanical excimer laser penetrating keratoplasty for perforated or predescemetal corneal ulcers.

Authors:  M Küchle; B Seitz; A Langenbucher; G O Naumann
Journal:  Ophthalmology       Date:  1999-11       Impact factor: 12.079

5.  Keratometric results of penetrating keratoplasty with the Hessburg-Barron and Hanna trephine systems using a standard double-running suture technique.

Authors:  S J Wiffen; L J Maguire; W M Bourne
Journal:  Cornea       Date:  1997-05       Impact factor: 2.651

6.  Clinical outcomes following penetrating keratoplasty using the Barron-Hessburg and Hanna corneal trephination systems.

Authors:  G A Wilbanks; S Cohen; M Chipman; D S Rootman
Journal:  Cornea       Date:  1996-11       Impact factor: 2.651

7.  Excimer laser "corneal shaping": a new technique for customized trephination in penetrating keratoplasty. First experimental results in rabbits.

Authors:  Klaus Schmitz; Wolfram Schreiber; Wolfgang Behrens-Baumann
Journal:  Graefes Arch Clin Exp Ophthalmol       Date:  2003-04-17       Impact factor: 3.117

8.  Comparison of astigmatism after penetrating keratoplasty by experienced cornea surgeons and cornea fellows.

Authors:  R H Gross; E J Poulsen; S Davitt; I R Schwab; M J Mannis
Journal:  Am J Ophthalmol       Date:  1997-05       Impact factor: 5.258

9.  [Immunologic transplant reactions after non-mechanical corneal trepanation with the excimer laser].

Authors:  B Seitz; A Langenbucher; M M Kus; M Küchle; G O Naumann
Journal:  Ophthalmologe       Date:  1998-09       Impact factor: 1.059

10.  Surgical advances and results of keratoconus.

Authors:  R C Troutman; R N Gaster
Journal:  Am J Ophthalmol       Date:  1980-08       Impact factor: 5.258

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.