Infrared thermographic SAR measurements of interstitial hyperthermia applicators: errors due to thermal conduction and convection.

M D Sherar1, A S Gladman, S R H Davidson, A C Easty, M L Joy.   

Abstract

Thermal conduction and convection were examined as sources of error in thermographically measured SAR patterns of an interstitial microwave hyperthermia applicator. Measurements were performed in a layered block of muscle-equivalent phantom material using an infrared thermographic technique with varying heating duration. There was a 52.7% reduction in maximum SAR and 75.5% increase in 50% iso-SAR contour area for a 60-s heating duration relative to a 10-s heating duration. A finite element model of heat transfer in an homogeneous medium was used to model conductive and convective heat transfer during the thermographic measurement. Thermal conduction artefacts were found to significantly distort thermographically measured SAR patterns. Convective cooling, which occurs when phantom layers are exposed for thermal image acquisition, was found to significantly affect the magnitude, but not the spatial distribution, of thermographically measured SAR patterns. Results from this investigation suggest that the thermal diffusion artefacts can be minimized if the duration of the applied power pulse is restricted to 10 s or less.

Entities:  

Mesh:

Year:  2004        PMID: 15277026     DOI: 10.1080/02656730410001668366

Source DB:  PubMed          Journal:  Int J Hyperthermia        ISSN: 0265-6736            Impact factor:   3.914


  8 in total

1.  Improved hyperthermia treatment control using SAR/temperature simulation and PRFS magnetic resonance thermal imaging.

Authors:  Zhen Li; Martin Vogel; Paolo F Maccarini; Vadim Stakhursky; Brian J Soher; Oana I Craciunescu; Shiva Das; Omar A Arabe; Williams T Joines; Paul R Stauffer
Journal:  Int J Hyperthermia       Date:  2010-11-11       Impact factor: 3.914

2.  An analytical solution for improved HIFU SAR estimation.

Authors:  C R Dillon; U Vyas; A Payne; D A Christensen; R B Roemer
Journal:  Phys Med Biol       Date:  2012-06-22       Impact factor: 3.609

3.  Experimental measurement of microwave ablation heating pattern and comparison to computer simulations.

Authors:  Garron Deshazer; Punit Prakash; Derek Merck; Dieter Haemmerich
Journal:  Int J Hyperthermia       Date:  2016-07-18       Impact factor: 3.914

4.  Predicting variation in subject thermal response during transcranial magnetic resonance guided focused ultrasound surgery: Comparison in seventeen subject datasets.

Authors:  Urvi Vyas; Pejman Ghanouni; Casey H Halpern; Jeff Elias; Kim Butts Pauly
Journal:  Med Phys       Date:  2016-09       Impact factor: 4.071

5.  In vitro and in vivo evaluations of increased effective beam width for heat deposition using a split focus high intensity ultrasound (HIFU) transducer.

Authors:  Pretesh R Patel; Alfred Luk; Amirk Durrani; Sergio Dromi; Julian Cuesta; Mary Angstadt; Matthew R Dreher; Bradford J Wood; Victor Frenkel
Journal:  Int J Hyperthermia       Date:  2008-11       Impact factor: 3.914

6.  Prospects for Millimeter-Wave Compliance Measurement Technologies.

Authors:  Leeor Alon; Sami Gabriel; Gene Young Cho; Ryan Brown; Cem M Deniz
Journal:  IEEE Antennas Propag Mag       Date:  2017-04-04       Impact factor: 2.593

7.  Microwave applicator for hyperthermia treatment on in vivo melanoma model.

Authors:  Paolo Togni; Jan Vrba; Luca Vannucci
Journal:  Med Biol Eng Comput       Date:  2009-12-24       Impact factor: 2.602

8.  Tumor selective hyperthermia induced by short-wave capacitively-coupled RF electric-fields.

Authors:  Mustafa Raoof; Brandon T Cisneros; Stuart J Corr; Flavio Palalon; Steven A Curley; Nadezhda V Koshkina
Journal:  PLoS One       Date:  2013-07-04       Impact factor: 3.240

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.