An evaluation of the heat and moisture exchange performance of four ventilator circuit filters.

C Jackson1, A R Webb.   

Abstract

OBJECTIVE: To compare the heat and moisture exchange efficiency of 4 commonly used ventilator circuit filters--DAR Hygroster (DHS), DAR Hygobac (DHC), Pall Ultipor BB50 (PUBC) and Intersurgical filtatherm (IFT).
DESIGN: Prospective randomized study.
SETTING: Intensive care unit of a university teaching hospital. PATIENTS AND PARTICIPANTS: 80 patients requiring post-operative mechanical ventilation were entered into the study. A total of 40 patients were studied after 1 h and 40 after 24 h. Due to technical errors 6 of the 1 h studies and 2 of the 24 h studies were excluded from analysis.
INTERVENTIONS: Mean wet and dry ceramic-platinum thermal probes were inserted in the inspiratory side of a device which separated inspiratory and expiratory gas flows on the patient side of the ventilator circuit filter. Humidification efficiency was calculated with reference to these wet and dry bulb temperatures. Tracheal temperature was measured by a similar thermal probe in the tracheal tube. MEASUREMENTS AND
RESULTS: Absolute humidity was greatest in group DHS (30.4 +/- 3.5 versus 22.9 +/- 1.5 mg/l at 1 h and 27.3 +/- 3.5 versus 20.7 +/- 2.4 mg/l at 24 h). The differences in absolute humidity reached statistical significance as follows: p = 0.0001 for DHS versus all others at 1 h and for DHS and DHC versus PUBC and IFT at 24 h.
CONCLUSION: Better humidification capabilities were obtained with groups DHS and DHC than PUBC and IFT.

Entities:  

Mesh:

Year:  1992        PMID: 1527255     DOI: 10.1007/bf01706470

Source DB:  PubMed          Journal:  Intensive Care Med        ISSN: 0342-4642            Impact factor:   17.440


  25 in total

1.  Mucous flow and ciliary activity in the trachea of healthy rats and rats exposed to respiratory irritant gases (SO2, H3N, HCHO): a functional and morphologic (light microscopic and electron microscopic) study, with special reference to technique.

Authors:  T DALHAMN
Journal:  Acta Physiol Scand Suppl       Date:  1956

2.  A comparative study of condenser humidifiers.

Authors:  A K Walker; D W Bethune
Journal:  Anaesthesia       Date:  1976-10       Impact factor: 6.955

3.  Endotracheal tube occlusion associated with the use of heat and moisture exchangers in the intensive care unit.

Authors:  I L Cohen; P F Weinberg; I A Fein; G S Rowinski
Journal:  Crit Care Med       Date:  1988-03       Impact factor: 7.598

Review 4.  Humidification in anaesthesia. A review of the present situation.

Authors:  J E Boys; T H Howells
Journal:  Br J Anaesth       Date:  1972-08       Impact factor: 9.166

5.  Temperature, humidity and mucus flow in the intubated trachea.

Authors:  A R Forbes
Journal:  Br J Anaesth       Date:  1974-01       Impact factor: 9.166

6.  Laboratory investigation of six artificial noses for use during endotracheal anesthesia.

Authors:  D B Weeks; F M Ramsey
Journal:  Anesth Analg       Date:  1983-08       Impact factor: 5.108

7.  Heat and moisture exchangers and vaporizing humidifiers in the intensive care unit.

Authors:  C Martin; G Perrin; M J Gevaudan; P Saux; F Gouin
Journal:  Chest       Date:  1990-01       Impact factor: 9.410

8.  Contamination control in long-term ventilation. A clinical study using a heat- and moisture-exchanging filter.

Authors:  J Gallagher; J E Strangeways; J Allt-Graham
Journal:  Anaesthesia       Date:  1987-05       Impact factor: 6.955

9.  The potential role of respiratory therapy equipment in cross infection. A study using a canine model for pneumonia.

Authors:  K L Christopher; L D Saravolatz; T L Bush; W A Conway
Journal:  Am Rev Respir Dis       Date:  1983-08

10.  Humidity and the anesthetized patient.

Authors:  J Chalon; C Patel; M Ali; S Ramanathan; L Capan; C K Tang; H Turndorf
Journal:  Anesthesiology       Date:  1979-03       Impact factor: 7.892

View more
  4 in total

1.  Factors associated with blocked tracheal tubes.

Authors:  F N Kapadia
Journal:  Intensive Care Med       Date:  2001-10       Impact factor: 17.440

2.  An evaluation of the microbial retention performance of three ventilator-circuit filters.

Authors:  J Holton; A R Webb
Journal:  Intensive Care Med       Date:  1994       Impact factor: 17.440

3.  Ventilator-associated pneumonia using a heated humidifier or a heat and moisture exchanger: a randomized controlled trial [ISRCTN88724583].

Authors:  Leonardo Lorente; María Lecuona; Alejandro Jiménez; María L Mora; Antonio Sierra
Journal:  Crit Care       Date:  2006       Impact factor: 9.097

4.  Evaluating humidity recovery efficiency of currently available heat and moisture exchangers: a respiratory system model study.

Authors:  Jeanette Janaina Jaber Lucato; Alexander Bernard Adams; Rogério Souza; Jamili Anbar Torquato; Carlos Roberto Ribeiro Carvalho; John J Marini
Journal:  Clinics (Sao Paulo)       Date:  2009       Impact factor: 2.365

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.