Literature DB >> 1526871

Source configuration and dose rates for the Selectron afterloading equipment for gynecologic applicators.

P W Grigsby1, J F Williamson, C A Perez.   

Abstract

This paper describes source train configurations of the Selectron LDR-6 that can be used to duplicate commonly used manually afterloadable source configurations for gynecological intracavitary brachytherapy. We demonstrate that the recommended source configuration using 2.5 mm 137Cs spheres is dosimetrically equivalent to its manually afterloaded counterparts using 3M 137Cs sources (20 mm overall length, 14 mm active length). Isodose distributions are in agreement to within 5% for the Fletcher-Suit-Delclos family of tandems, colpostats, and vaginal cylinders. Our data also demonstrate that the isotropic point source model used by most computerized treatment planning systems is a good approximation to source trains of filtered, spherical sources interspersed with solid steel spacers. We conclude that the use of 2.5 mm 137Cs sources in the Selectron remote afterloading system can accurately reproduce the isodose distribution achieved with linear 137Cs manually afterloaded sources.

Mesh:

Substances:

Year:  1992        PMID: 1526871     DOI: 10.1016/0360-3016(92)90688-e

Source DB:  PubMed          Journal:  Int J Radiat Oncol Biol Phys        ISSN: 0360-3016            Impact factor:   7.038


  4 in total

1.  Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation.

Authors:  Somayeh Gholami; Hamid Reza Mirzaei; Ali Jabbary Arfaee; Ramin Jaberi; Hassan Ali Nedaie; Seied Rabi Mahdavi; Eftekhar Rajab Bolookat; Ali S Meigooni
Journal:  Rep Pract Oncol Radiother       Date:  2016-07-18

2.  Independent evaluation of an in-house brachytherapy treatment planning system using simulation, measurement and calculation methods.

Authors:  M A Mosleh Shirazi; R Faghihi; Z Siavashpour; H A Nedaie; S Mehdizadeh; S Sina
Journal:  J Appl Clin Med Phys       Date:  2012-03-08       Impact factor: 2.102

3.  Perturbation of TG-43 parameters of the brachytherapy sources under insufficient scattering materials.

Authors:  Mehdi Zehtabian; Sedigheh Sina; Reza Faghihi; Ali Meigooni
Journal:  J Appl Clin Med Phys       Date:  2013-05-06       Impact factor: 2.102

4.  Impact of the vaginal applicator and dummy pellets on the dosimetry parameters of Cs-137 brachytherapy source.

Authors:  Sedigheh Sina; Reza Faghihi; Ali S Meigooni; Simin Mehdizadeh; M Amin Mosleh Shirazi; Mehdi Zehtabian
Journal:  J Appl Clin Med Phys       Date:  2011-05-19       Impact factor: 2.243

  4 in total