Literature DB >> 15260418

Acetabular cementing technique in THA--flanged versus unflanged cups, cadaver experiments.

Dominik Parsch1, Caroline Diehm, Sven Schneider, Andrew New, Steffen J Breusch.   

Abstract

BACKGROUND: There are few studies on the effect of acetabular cup design on cement penetration.
MATERIAL AND METHODS: We evaluated the effects of an acetabular flange on cement pressurization and cement penetration in 12 cadavers. Flanged or unflanged cups were implanted in paired human acetabula with simulated intraosseous bleeding pressure but without cement pressurization before insertion of the cup. Three pressure transducers were used to record intra-acetabular peak and average pressures during cup insertion. Following implantation, the whole specimens were AP-radiographed and standardized sections through the acetabula were microradiographed to evaluate cement penetration.
RESULTS: Flanged cups produced greater intra-acetabular peak pressures than unflanged cups, but did not increase the average intra-acetabular pressure. Cement penetration did not differ significantly between the two groups.
INTERPRETATION: Our findings do not support the use of flanged cups as the sole means of cement pressurization in the acetabulum.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15260418     DOI: 10.1080/00016470410001196

Source DB:  PubMed          Journal:  Acta Orthop Scand        ISSN: 0001-6470


  6 in total

1.  Basic science considerations in primary total hip replacement arthroplasty.

Authors:  Saqeb B Mirza; Douglas G Dunlop; Sukhmeet S Panesar; Syed G Naqvi; Shafat Gangoo; Saif Salih
Journal:  Open Orthop J       Date:  2010-05-11

2.  The use of a "rim cutter" device and a flanged cup for improving the mantle of the acetabular component of a cemented Exeter total hip arthroplasty.

Authors:  Jenaro Ángel Fernández-Valencia; Xavier Gallart; Guillem Bori; Juan Miguel Rodríguez-Roiz; Andreu Combalia
Journal:  Eur J Orthop Surg Traumatol       Date:  2016-08-22

3.  Comparison of flanged and unflanged acetabular cup design. An experimental study using ceramic and cadaveric acetabuli.

Authors:  Mette Ørskov; Saba Abdulghani; Ian McCarthy; Kjeld Søballe; Gunnar Flivik
Journal:  Acta Orthop       Date:  2010-10       Impact factor: 3.717

4.  Comparison of cement pressurisation in flanged and unflanged acetabular cups.

Authors:  Rajarshi Bhattacharya; Fahad G Attar; Sarah Green; Andrew Port
Journal:  J Orthop Surg Res       Date:  2012-02-03       Impact factor: 2.359

5.  The addition of a flange does not improve the pressure generated during cemented acetabular cup implantation.

Authors:  Alexander T Boote; David J Deehan; Kenneth S Rankin; David C Swailes; Philip J Hyde
Journal:  J Biomed Mater Res B Appl Biomater       Date:  2022-06-03       Impact factor: 3.405

6.  What is the Optimum Pattern of Pressurisation to Gain Maximum Penetration of Methylmethacrylate Cement into the Reamed Acetabulum?

Authors:  Brett Rocos; Michael R Whitehouse
Journal:  Cureus       Date:  2020-01-14
  6 in total