[Expandable cages: biomechanical comparison of different cages for ventral spondylodesis in the thoracolumbar spine].

C Khodadadyan-Klostermann1, J Schaefer, Ph Schleicher, R Pflugmacher, T Eindorf, N P Haas, F Kandziora.   

Abstract

INTRODUCTION: Due to a recent increase in the commercial availability of expandable cages for vertebral body replacement, this study was designed to gain more information about their biomechanical properties. All three expandable cages evaluated in this study are approved for clinical use, but little knowledge about their biomechanical properties exists.
MATERIAL AND METHODS: Human thoracolumbar spines (T11 to L3) ( n=32) were tested in flexion, extension, axial rotation, and lateral bending with a nondestructive stiffness method. Three-dimensional displacement was measured using an optical measurement system. All motion segments were tested intact. After L1 corporectomy, cages were implanted and the following groups ( n=8 each) were tested: (1) meshed titanium cage (nonexpandable cage, DePuy Acromed), (2) X-tenz (expandable cage, DePuy Acromed), (3) Synex (expandable Cage, Synthes), and (4) VBR (expandable cage, Ulrich). Finally, posterior stabilization and posterior-anterior stabilization, both using USS (Synthes), and anterior plating (LCDCP, Synthes) was applied. The mean apparent stiffness values, ranges of motion, and neutral and elastic zones were calculated from the corresponding load/displacement curves.
RESULTS: No significant differences were found between the in vitro biomechanical properties of expandable and nonexpandable cages. Compared to the intact motion segment, isolated anterior stabilization using cages and anterior plating significantly decreased stiffness and increased range of motion in all directions. Additional posterior stabilization significantly increased stiffness and decreased range of motion in all directions compared to the intact motion segment. Combined anterior-posterior stabilization demonstrated the greatest stiffness results.
CONCLUSION: Design variations of expandable cages for vertebral body replacement do not show any significant effect on the biomechanical results. There was no significant difference found, between the biomechanical properties of expandable and non-expandable cages. After corporectomy, isolated implantation of expandable cages plus anterior plating was not able to restore normal stability of the motion segment. As a consequence, isolated anterior stabilization using cages plus LCDCP should not be used for vertebral body replacement in the thoraco-lumbar spine.

Entities:  

Mesh:

Year:  2004        PMID: 15258751     DOI: 10.1007/s00104-003-0786-4

Source DB:  PubMed          Journal:  Chirurg        ISSN: 0009-4722            Impact factor:   0.955


  25 in total

1.  Three-dimensional motion analysis with Synex. Comparative biomechanical test series with a new vertebral body replacement for the thoracolumbar spine.

Authors:  C Knop; U Lange; L Bastian; M Blauth
Journal:  Eur Spine J       Date:  2000-12       Impact factor: 3.134

2.  Testing criteria for spinal implants: recommendations for the standardization of in vitro stability testing of spinal implants.

Authors:  H J Wilke; K Wenger; L Claes
Journal:  Eur Spine J       Date:  1998       Impact factor: 3.134

Review 3.  [New developments in surgery of the injured spine].

Authors:  M Blauth; C Knop; L Bastian; P Lobenhoffer
Journal:  Orthopade       Date:  1997-05       Impact factor: 1.087

4.  Flexible device for vertebral body replacement.

Authors:  J A Main; M E Wells; D M Spengler; A M Strauss; T S Keller
Journal:  J Biomed Eng       Date:  1989-03

5.  Anterior interbody lumbar spine fusion. Analysis of Mayo Clinic series.

Authors:  R N Stauffer; M B Coventry
Journal:  J Bone Joint Surg Am       Date:  1972-06       Impact factor: 5.284

6.  Stability of the lumbar spine and method of instrumentation.

Authors:  L Vanden Berghe; H Mehdian; A J Lee; C R Weatherley
Journal:  Acta Orthop Belg       Date:  1993       Impact factor: 0.500

7.  Anterior interbody fusion of the lumbar spine. Indications and results in 200 cases.

Authors:  S Sacks
Journal:  Clin Orthop Relat Res       Date:  1966 Jan-Feb       Impact factor: 4.176

8.  Biomechanical comparison of cervical spine interbody fusion cages.

Authors:  F Kandziora; R Pflugmacher; J Schäfer; C Born; G Duda; N P Haas; T Mittlmeier
Journal:  Spine (Phila Pa 1976)       Date:  2001-09-01       Impact factor: 3.468

9.  Anterior decompression and stabilization of thoracolumbar burst fractures by the use of the Slot-Zielke device.

Authors:  H D Been
Journal:  Spine (Phila Pa 1976)       Date:  1991-01       Impact factor: 3.468

10.  Morbidity at bone graft donor sites.

Authors:  E M Younger; M W Chapman
Journal:  J Orthop Trauma       Date:  1989       Impact factor: 2.512

View more
  6 in total

1.  [History of vertebral body replacement].

Authors:  T Tarhan; D Froemel; M Rickert; M Rauschmann; C Fleege
Journal:  Unfallchirurg       Date:  2015-12       Impact factor: 1.000

2.  Biomechanical analysis of a new expandable vertebral body replacement combined with a new polyaxial antero-lateral plate and/or pedicle screws and rods.

Authors:  Benjamin Ulmar; Stefanie Erhart; Stefan Unger; Kuno Weise; Werner Schmoelz
Journal:  Eur Spine J       Date:  2011-10-18       Impact factor: 3.134

3.  Late collapse osteoporotic vertebral fracture in an elderly patient with neurological compromise.

Authors:  D Ruiz Picazo; J Ramírez Villaescusa; E Portero Martínez; F Doñate Pérez
Journal:  Eur Spine J       Date:  2013-06-19       Impact factor: 3.134

4.  Osteoporotic burst fracture-clinical, radiological and functional outcome of three-column reconstruction using single posterior approach (Instrumentation, Corpectomy, Arthroscope Assisted Transpedicular Decompression and Mesh Cage).

Authors:  Ghanshyam Kakadiya; Viraj Gandbhir; Yogesh Soni; Kushal Gohil; Akash Shakya
Journal:  N Am Spine Soc J       Date:  2020-05-29

5.  Redundant publications in surgery: a threat to patient safety?

Authors:  Philip F Stahel; Pierre-Alain Clavien; Wade R Smith; Ernest E Moore
Journal:  Patient Saf Surg       Date:  2008-03-19

6.  Functional Outcomes of Nerve Root Sparing Posterior Corpectomy in Lumbar Vertebral Burst Fractures.

Authors:  Abhinandan Reddy Mallepally; Nandan Marathe; Abhinav Kumar Shrivastava; Vikas Tandon; Harvinder Singh Chhabra
Journal:  Global Spine J       Date:  2021-01-25
  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.