Optimization of atrial defibrillation with a dual-coil, active pectoral lead system.

Eric J Rashba1, Stephen R Shorofsky, Robert W Peters, Michael R Gold.   

Abstract

INTRODUCTION: Atrial defibrillation can be achieved with standard implantable cardioverter defibrillator (ICD) leads, but the optimal shocking configuration is unknown. The objective of this prospective study was to compare atrial defibrillation thresholds (DFTs) with three shocking configurations that are available with standard ICD leads. METHODS AND
RESULTS: This study was a prospective, randomized, paired comparison of shocking configurations on atrial DFTs in 58 patients. The lead system evaluated was a transvenous defibrillation lead with coils in the superior vena cava (SVC) and right ventricular apex (RV) and a left pectoral pulse generator emulator (Can). In the first 33 patients, atrial DFT was measured with the ventricular triad (RV --> SVC + Can) and unipolar (RV --> Can) shocking pathways. In the next 25 patients, atrial DFT was measured with the ventricular triad and the proximal triad (SVC --> RV + Can) configurations. Delivered energy at DFT was significantly lower with the ventricular triad compared to the unipolar configuration (4.7 +/- 3.7 J vs 10.1 +/- 9.5 J, P < 0.001). Peak voltage and shock impedance also were significantly reduced (P < 0.001). There was no significant difference in DFT energy when the ventricular triad and proximal triad shocking configurations were compared (3.6 +/- 3.0 J vs 3.4 +/- 2.9 J for ventricular and proximal triad, respectively, P = NS). Although shock impedance was reduced by 13% with the proximal triad (P < 0.001), this effect was offset by an increased current requirement (10%).
CONCLUSION: The ventricular triad is equivalent or superior to other possible shocking pathways for atrial defibrillation afforded by a dual-coil, active pectoral lead system. Because the ventricular triad is also the most efficacious shocking pathway for ventricular defibrillation, this pathway should be preferred for combined atrial and ventricular defibrillators.

Entities:  

Mesh:

Year:  2004        PMID: 15250864     DOI: 10.1046/j.1540-8167.2004.03684.x

Source DB:  PubMed          Journal:  J Cardiovasc Electrophysiol        ISSN: 1045-3873


  4 in total

Review 1.  Dual- versus single-coil implantable defibrillator leads: review of the literature.

Authors:  Jörg Neuzner; Jörg Carlsson
Journal:  Clin Res Cardiol       Date:  2012-01-10       Impact factor: 5.460

2.  Extended charge banking model of dual path shocks for implantable cardioverter defibrillators.

Authors:  Derek J Dosdall; James D Sweeney
Journal:  Biomed Eng Online       Date:  2008-08-01       Impact factor: 2.819

3.  When, how, and why should sinus rhythm be restored in patients with persistent atrial fibrillation?

Authors:  Kent Stephenson; Cory M Tschabrunn; Sujethra Vasu; Eric J Rashba
Journal:  Curr Treat Options Cardiovasc Med       Date:  2007-10

4.  Low efficacy of cardioversion of persistent atrial fibrillation with the implantable cardioverter-defibrillator.

Authors:  I Limantoro; K Vernooy; B Weijs; R Pisters; L Debie; H J Crijns; Y Blaauw
Journal:  Neth Heart J       Date:  2013-12       Impact factor: 2.380

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.