Patient alert in implantable cardioverter defibrillators: toy or tool?

Ruediger Becker1, Jutta Ruf-Richter, Julia C Senges-Becker, Alexander Bauer, Slawomir Weretka, Frederik Voss, Hugo A Katus, Wolfgang Schoels.   

Abstract

OBJECTIVES: The purpose of this study was to analyze the utility of patient-alert features in implantable cardioverter defibrillators (ICDs).
BACKGROUND: Various alert features producing acoustic warning signals have been implemented in newer generation ICDs, but their role in early detection of system-related complications has not been systematically evaluated.
METHODS: In 240 patients implanted with Medtronic ICD devices, the following alert features were routinely activated: pacing lead impedance <200 or >2,000 Omega, high-voltage lead impedance <10 or >200 Omega, low battery voltage (elective replacement indicator), long charge time (>18 s), >3 shocks delivered per episode, and all therapies in a zone delivered. Alert events occurring during follow-up were assessed in relation to actual findings (hospital charts, chest X-rays, ICD printouts including sensing/pacing/defibrillation threshold tests, episode data) to determine incidence, sensitivity, and specificity of the alert function.
RESULTS: During 12.2 +/- 8.9 months, 24 alert events occurred in the 240 patients (pacing lead impedance, n = 4; high-voltage lead impedance, n = 7; low battery voltage, n = 1; >3 shocks, n = 6; all therapies, n = 6). A total of 22 serious complications (necessitating reprogramming or device/lead replacement) were observed, 14 of which were primarily identified through a patient alert (lead fracture, n = 11; connector defect, n = 1; T-wave oversensing, n = 1; battery depletion, n = 1). This reflects a sensitivity of 64% and a specificity of 96% of the alert function for serious complications. With 14 of 24 patient alerts being caused by serious complications, the positive predictive value reached 58%.
CONCLUSIONS: Patient-alert features are a useful additional tool facilitating early detection of serious ICD complications, but they do not substitute for regular ICD follow-up, because of their low sensitivity.

Entities:  

Mesh:

Year:  2004        PMID: 15234415     DOI: 10.1016/j.jacc.2004.03.051

Source DB:  PubMed          Journal:  J Am Coll Cardiol        ISSN: 0735-1097            Impact factor:   24.094


  9 in total

1.  Remote monitoring in implantable defibrillator therapy.

Authors:  D A M J Theuns; L S Jordaens
Journal:  Neth Heart J       Date:  2008-02       Impact factor: 2.380

2.  Can we predict and prevent adverse events related to high-voltage implantable cardioverter defibrillator lead failure?

Authors:  Renato Pietro Ricci; Carlo Pignalberi; Barbara Magris; Stefano Aquilani; Vito Altamura; Loredana Morichelli; Antonio Porfili; Laura Quarta; Fabio Saputo; Massimo Santini
Journal:  J Interv Card Electrophysiol       Date:  2011-09-01       Impact factor: 1.900

Review 3.  Troubleshooting implantable cardioverter-defibrillators: an overview for physicians who are not electrophysiologists.

Authors:  Mohammad Saeed
Journal:  Tex Heart Inst J       Date:  2011

Review 4.  [Home Monitoring with implantable ICD--a diagnostic innovation?].

Authors:  W Jung; R Birkemeyer
Journal:  Herzschrittmacherther Elektrophysiol       Date:  2005-09

5.  What evidence do we have to replace in-hospital implantable cardioverter defibrillator follow-up?

Authors:  P Brugada
Journal:  Clin Res Cardiol       Date:  2006       Impact factor: 5.460

Review 6.  Telecardiology and remote monitoring of implanted electrical devices: the potential for fresh clinical care perspectives.

Authors:  Giuseppe Boriani; Igor Diemberger; Cristian Martignani; Mauro Biffi; Cinzia Valzania; Matteo Bertini; Giulia Domenichini; Davide Saporito; Matteo Ziacchi; Angelo Branzi
Journal:  J Gen Intern Med       Date:  2008-01       Impact factor: 5.128

7.  Adding wisdom to 'smart' bioelectronic systems: a design framework for physiologic control including practical examples.

Authors:  Aysegul Gunduz; Enrico Opri; Ro'ee Gilron; Vaclav Kremen; Gregory Worrell; Phil Starr; Kent Leyde; Timothy Denison
Journal:  Bioelectron Med (Lond)       Date:  2019-05-30

8.  Design of the evolution of management strategies of heart failure patients with implantable defibrillators (EVOLVO) study to assess the ability of remote monitoring to treat and triage patients more effectively.

Authors:  Maurizio Marzegalli; Maurizio Landolina; Maurizio Lunati; Giovanni B Perego; Alessia Pappone; Giuseppe Guenzati; Carlo Campana; Maria Frigerio; Gianfranco Parati; Antonio Curnis; Irene Colangelo; Sergio Valsecchi
Journal:  Trials       Date:  2009-06-18       Impact factor: 2.279

9.  Patient alerting features in implantable defibrillators.

Authors:  Dirk Vollmann; Markus Zabel; Johnson Francis
Journal:  Indian Pacing Electrophysiol J       Date:  2008-02-01
  9 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.