| Literature DB >> 15234172 |
Abstract
A linear array imaging system with coded excitation is considered, where the proposed excitation/compression scheme maximizes the signal-to-noise ratio (SNR) and minimizes sidelobes at the output of the compression filter. A pulse with linear frequency modulation (LFM) is used for coded excitation. The excitation/compression scheme is based on the fast digital mismatched filtering. The parameter optimization of the excitation/compression scheme includes (i) choice of an optimal filtering function for the mismatched filtering; (ii) choice of an optimal window function for tapering of the chirp amplitude; (iii) optimization of a chirp-to-transducer bandwidth ratio; (iv) choice of an appropriate n-bit quantizer. The simulation results show that the excitation/compression scheme can be implemented as a Dolph-Chebyshev filter including amplitude tapering of the chirp with a Lanczos window. An example of such an optimized system is given where the chirp bandwidth is chosen to be 2.5 times the transducer bandwidth and equals 6 MHz: The sidelobes are suppressed to -80 dB, for a central frequency of 4 MHz, and to -94 dB, for a central frequency of 8 MHz. The corresponding improvement of the SNR is 18 and 21 dB, respectively, when compared to a conventional short pulse imaging system. Simulation of B-mode images demonstrates the advantage of coded excitation systems of detecting regions with low contrast.Year: 2004 PMID: 15234172 DOI: 10.1016/j.ultras.2004.02.020
Source DB: PubMed Journal: Ultrasonics ISSN: 0041-624X Impact factor: 2.890