Abstract

OBJECTIVE: To evaluate the accuracy of automated digital blood pressure monitoring devices and operators in the community. Also, we tested the effects of a simple education program, and looked for arm-arm differences.
DESIGN: Subjects who had bought their own automated digital blood pressure monitor were recruited via an advertisement in the local newspaper. On arrival, they were asked to record their blood pressure exactly as they would at home. The investigator noted any technique deficiencies then corrected them. Blood pressures were then recorded by the investigator and the subject, on opposite arms, simultaneously, and repeated with the arms switched. Finally, subjects recorded their blood pressure again. The subjects' readings were compared to the average of monitor and mercury readings using Bland-Altman methods.
RESULTS: A total of 80 subjects were tested. Before educating, subjects' systolic blood pressure (SBP) readings were +5.8+/-6.4 (standard deviation) mmHg greater than the mean of all readings, and diastolic blood pressure (DBP) were +1.3+/-4.0 mmHg; after educating they were +1.3+/-4.0 and -1.3+/-2.7 respectively. The monitors, as a group, were accurate, and met British Hypertension Society and AAMI highest standards. We found no differences among monitors that had been validated (n=26) and those that had not. There were differences between the arms: 5.3+/-5.2 mmHg for SBP and 3.4+/-3.3 mmHg for DBP. Most patients had never been informed by anyone of proper blood pressure measuring techniques.
CONCLUSIONS: We conclude that home blood pressure measurement, as practiced in our community, is prone to error, mostly due to mistakes by the operator. These can easily be corrected, so that readings become more accurate. Attention should be paid to arm-arm differences.

Entities:  

Mesh:

Year:  2004        PMID: 15199308     DOI: 10.1097/01.mbp.0000130224.18777.ce

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  8 in total

1.  Performance and persistence of a blood pressure self-management intervention: telemonitoring and self-management in hypertension (TASMINH2) trial.

Authors:  E P Bray; M I Jones; M Banting; S Greenfield; F D R Hobbs; P Little; B Williams; R J Mcmanus
Journal:  J Hum Hypertens       Date:  2015-01-08       Impact factor: 3.012

2.  Public health in an era of personal health records: opportunities for innovation and new partnerships.

Authors:  Jason Bonander; Suzanne Gates
Journal:  J Med Internet Res       Date:  2010-08-10       Impact factor: 5.428

3.  Applications of electronic health information in public health: uses, opportunities & barriers.

Authors:  Alan Tomines; Heather Readhead; Adam Readhead; Steven Teutsch
Journal:  EGEMS (Wash DC)       Date:  2013-10-28

4.  How Accurate Are Home Blood Pressure Devices in Use? A Cross-Sectional Study.

Authors:  Marcel Ruzicka; Ayub Akbari; Eva Bruketa; Jeanne Françoise Kayibanda; Claude Baril; Swapnil Hiremath
Journal:  PLoS One       Date:  2016-06-01       Impact factor: 3.240

5.  Impact of cuff positioning on blood pressure measurement accuracy: may a specially designed cuff make a difference?

Authors:  Grzegorz Bilo; Oscar Sala; Carlotta Perego; Andrea Faini; Lan Gao; Anna Głuszewska; Juan Eugenio Ochoa; Dario Pellegrini; Laura Maria Lonati; Gianfranco Parati
Journal:  Hypertens Res       Date:  2017-01-12       Impact factor: 3.872

Review 6.  The Impact of COVID-19 on Patients With ADPKD.

Authors:  Meherzad Kutky; Erin Cross; Darin J Treleaven; Ahsan Alam; Matthew B Lanktree
Journal:  Can J Kidney Health Dis       Date:  2021-11-10

7.  Patient self-monitoring of blood pressure and self-titration of medication in primary care: the TASMINH2 trial qualitative study of health professionals' experiences.

Authors:  Miren I Jones; Sheila M Greenfield; Emma P Bray; Fd Richard Hobbs; Roger Holder; Paul Little; Jonathan Mant; Bryan Williams; Richard J McManus
Journal:  Br J Gen Pract       Date:  2013-06       Impact factor: 5.386

8.  Accuracy of blood-pressure monitors owned by patients with hypertension (ACCU-RATE study): a cross-sectional, observational study in central England.

Authors:  James A Hodgkinson; Mei-Man Lee; Siobhan Milner; Peter Bradburn; Richard Stevens; Fd Richard Hobbs; Constantinos Koshiaris; Sabrina Grant; Jonathan Mant; Richard J McManus
Journal:  Br J Gen Pract       Date:  2020-07-30       Impact factor: 5.386

  8 in total