Literature DB >> 15193877

Comparison of metal release from various metallic biomaterials in vitro.

Yoshimitsu Okazaki1, Emiko Gotoh.   

Abstract

To investigate the metal release of each base and alloying elements in vitro, SUS316L stainless steel, Co-Cr-Mo casting alloy, commercially pure Ti grade 2, and Ti-6Al-4V, V-free Ti-6Al-7Nb and Ti-15Zr-4Nb-4Ta alloys were immersed in various solutions, namely, alpha-medium, PBS(-), calf serum, 0.9% NaCl, artificial saliva, 1.2 mass% L-cysteine, 1 mass% lactic acid and 0.01 mass% HCl for 7d. The difference in the quantity of Co released from the Co-Cr-Mo casting alloy was relatively small in all the solutions. The quantities of Ti released into alpha-medium, PBS(-), calf serum, 0.9% NaCl and artificial saliva were much lower than those released into 1.2% L-cysteine, 1% lactic acid and 0.01% HCl. The quantity of Fe released from SUS316L stainless steel decreased linearly with increasing pH. On the other hand, the quantity of Ti released from Ti materials increased with decreasing pH, and it markedly attenuated at pHs of approximately 4 and higher. The quantity of Ni released from stainless steel gradually decreased with increasing pH. The quantities of Al released from the Ti-6Al-4V and Ti-6Al-7Nb alloys gradually decreased with increasing pH. A small V release was observed in calf serum, PBS(-), artificial saliva, 1% lactic acid, 1.2% l-cysteine and 0.01% HCl. The quantity of Ti released from the Ti-15Zr-4Nb-4Ta alloy was smaller than those released from the Ti-6Al-4V and Ti-6Al-7Nb alloys in all the solutions. In particular, it was approximately 30% or smaller in 1% lactic acid, 1.2% L-cysteine and 0.01% HCl. The quantity of (Zr + Nb + Ta) released was also considerably lower than that of (Al + Nb) or (Al + V) released. Therefore, the Ti-15Zr-4Nb-4Ta alloy with its low metal release in vitro is considered advantageous for long-term implants. Copyright 2004 Elsevier Ltd.

Entities:  

Mesh:

Substances:

Year:  2005        PMID: 15193877     DOI: 10.1016/j.biomaterials.2004.02.005

Source DB:  PubMed          Journal:  Biomaterials        ISSN: 0142-9612            Impact factor:   12.479


  60 in total

1.  Mechanical alloying of biocompatible Co-28Cr-6Mo alloy.

Authors:  F Sánchez-De Jesús; A M Bolarín-Miró; G Torres-Villaseñor; C A Cortés-Escobedo; J A Betancourt-Cantera
Journal:  J Mater Sci Mater Med       Date:  2010-04-03       Impact factor: 3.896

2.  Grit blasting of medical stainless steel: implications on its corrosion behavior, ion release and biocompatibility.

Authors:  J C Galván; L Saldaña; M Multigner; A Calzado-Martín; M Larrea; C Serra; N Vilaboa; J L González-Carrasco
Journal:  J Mater Sci Mater Med       Date:  2012-01-22       Impact factor: 3.896

3.  Investigating the immunologic effects of CoCr nanoparticles.

Authors:  Bamikole Ogunwale; Andreas Schmidt-Ott; R M Dominic Meek; James M Brewer
Journal:  Clin Orthop Relat Res       Date:  2009-06-30       Impact factor: 4.176

4.  Deposition of TiN films on Co-Cr for improving mechanical properties and biocompatibility using reactive DC sputtering.

Authors:  Vuong-Hung Pham; Se-Won Yook; Eun-Jung Lee; Yuanlong Li; Gyuran Jeon; Jung-Joong Lee; Hyoun-Ee Kim; Young-Hag Koh
Journal:  J Mater Sci Mater Med       Date:  2011-08-13       Impact factor: 3.896

5.  Stability of plasma electrolytic oxidation coating on titanium in artificial saliva.

Authors:  E Matykina; R Arrabal; M Mohedano; A Pardo; M C Merino; E Rivero
Journal:  J Mater Sci Mater Med       Date:  2012-10-17       Impact factor: 3.896

6.  Influence of noble metals alloying additions on the corrosion behaviour of titanium in a fluoride-containing environment.

Authors:  F Rosalbino; S Delsante; G Borzone; G Scavino
Journal:  J Mater Sci Mater Med       Date:  2012-03-02       Impact factor: 3.896

7.  Development of superlattice CrN/NbN coatings for joint replacements deposited by high power impulse magnetron sputtering.

Authors:  Papken Ehiasarian Hovsepian; Arutiun Papken Ehiasarian; Yashodhan Purandare; Arunprabhu Arunachalam Sugumaran; Tim Marriott; Imran Khan
Journal:  J Mater Sci Mater Med       Date:  2016-08-29       Impact factor: 3.896

8.  Novel Zinc / Tungsten Carbide Nanocomposite as Bioabsorbable Implant.

Authors:  Zeyi Guan; Chase S Linsley; Injoo Hwang; Gongcheng Yao; Benjamin M Wu; Xiaochun Li
Journal:  Mater Lett       Date:  2019-12-28       Impact factor: 3.423

Review 9.  Are clinical findings of systemic titanium dispersion following implantation explained by available in vitro evidence? An evidence-based analysis.

Authors:  Justin Paul Curtin; Minji Wang
Journal:  J Biol Inorg Chem       Date:  2017-05-17       Impact factor: 3.358

10.  Electrochemical characterization of AISI 316L stainless steel in contact with simulated body fluid under infection conditions.

Authors:  Danián Alejandro López; Alicia Durán; Silvia Marcela Ceré
Journal:  J Mater Sci Mater Med       Date:  2007-11-13       Impact factor: 3.896