Literature DB >> 15186243

Torsional fatigue and endurance limit of a size 30.06 ProFile rotary instrument.

S Best1, P Watson, R Pilliar, G G K Kulkarni, G Yared.   

Abstract

AIM: To evaluate the torsional cyclic fatigue characteristics and specifically the endurance limit (EL) of a nickel-titanium rotary instrument.
METHODOLOGY: Size 30.06 taper ProFile instruments were evaluated. The equipment was assembled according to the ANSI/ADA Specification No. 28. The motor was programmed to repeatedly rotate to a selected deflection angle (DA) and then return to zero (cycle). Testing started at 200 degrees and was continued at decreasing angles until 10(6) cycles were reached without instrument fracture. Ten instruments were tested at each DA. The mean log number of cycles to fracture and standard deviation were determined for each DA at which fracture occurred. The DA at which 10(6) cycles were reached without instrument fracture corresponded by definition to the EL. Analysis of variance and pairwise comparisons using Duncan's multiple range test were performed to detect significant differences among the mean log number of cycles of the different DA. Significance was determined at the 95% confidence level.
RESULTS: Instruments cycled at larger DA consistently demonstrated fewer cycles to fracture than those cycled at smaller DA. The differences among the mean log number of cycles of the different DA were statistically significant (P < 0.001). Cycles of 10(6) were completed without instrument fracture at 2.5 degrees.
CONCLUSIONS: A torsional fatigue profile was generated for a specific NiTi rotary instrument. The EL was 2.5 degrees.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15186243     DOI: 10.1111/j.1365-2591.2004.00814.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  5 in total

1.  Deformations and cyclic fatigue resistance of nickel-titanium instruments inside a sequence.

Authors:  Gianluca Gambarini; Gianluca Plotino; Lucila Piasecki; Dina Al-Sudani; Luca Testarelli; Gianpaolo Sannino
Journal:  Ann Stomatol (Roma)       Date:  2015-05-18

2.  Canal preparation with nickel-titanium or stainless steel instruments without the risk of instrument fracture: preliminary observations.

Authors:  Ghassan Yared
Journal:  Restor Dent Endod       Date:  2014-11-13

3.  Microscopic Features of Fractured Fragment of Nickel-Titanium Endodontic Instruments by Two Different Modes of Torsional Loading.

Authors:  Ibrahim H Abu-Tahun; Sang Won Kwak; Jung-Hong Ha; Hyeon-Cheol Kim
Journal:  Scanning       Date:  2018-02-21       Impact factor: 1.932

4.  Evaluation of dynamic and static torsional resistances of nickel-titanium rotary instruments.

Authors:  Ibrahim H Abu-Tahun; Jung-Hong Ha; Sang Won Kwak; Hyeon-Cheol Kim
Journal:  J Dent Sci       Date:  2018-02-01       Impact factor: 2.080

5.  Resistance to torsional failure and cyclic fatigue resistance of ProTaper Next, WaveOne, and Mtwo files in continuous and reciprocating motion: An in vitro study.

Authors:  N O Varghese; Rajesh Pillai; U-Nu Sujathen; Shan Sainudeen; Abe Antony; Sinju Paul
Journal:  J Conserv Dent       Date:  2016 May-Jun
  5 in total