Literature DB >> 15157292

Endovascular occlusion of wide-necked aneurysms with a new intracranial microstent (Neuroform) and detachable coils.

Ronald P Benitez1, Marco T Silva, Jack Klem, Erol Veznedaroglu, Robert H Rosenwasser.   

Abstract

OBJECTIVE: The long-term durability of the endovascular occlusion of cerebral aneurysms is one of the major factors limiting the more widespread use of this technique. Long-term occlusion of wide-necked aneurysms has improved with new assistive devices that seem to improve aneurysm occlusion while protecting the parent vessel. We report the use of a new intracranial stent--the Neuroform microstent--in the treatment of patients with wide-necked cerebral aneurysms.
METHODS: Patients identified as harboring wide-necked intracranial aneurysms were evaluated for stent-assisted coiling. After appropriate anticoagulation was performed, depending on whether the aneurysm was ruptured or unruptured, the Neuroform stent was delivered across the neck of the aneurysm and deployed with a coil pusher. After stent placement, standard coil occlusion of the aneurysm was achieved in the majority of cases.
RESULTS: Fifty-six patients were identified as having wide-necked intracranial aneurysms suitable for stent-assisted coiling. A total of 49 aneurysms in 48 patients were treated with this procedure. In eight cases, stent deployment failed. Forty-one of the aneurysms were initially stented, followed by coil placement. Six aneurysms were stented only, and one aneurysm was initially coiled, followed by stent placement. There were five deaths (8.9%), one of which occurred secondary to a stroke after the procedure (1.8%). Four patients (7%) experienced thromboembolic events, three of which were considered to have been secondary to the procedure (5.3%). In addition, there were two femoral pseudoaneurysms. The overall complication rate was 10.7%. Five patients were available for follow-up angiographic evaluation, and their cases are discussed.
CONCLUSION: Intracranial stenting may overcome important technical limitations in current endovascular therapy by improving the occlusion of wide-necked aneurysms while protecting the parent vessel.

Entities:  

Mesh:

Year:  2004        PMID: 15157292     DOI: 10.1227/01.neu.0000124484.87635.cd

Source DB:  PubMed          Journal:  Neurosurgery        ISSN: 0148-396X            Impact factor:   4.654


  115 in total

Review 1.  Review of 2 decades of aneurysm-recurrence literature, part 1: reducing recurrence after endovascular coiling.

Authors:  E Crobeddu; G Lanzino; D F Kallmes; H J Cloft
Journal:  AJNR Am J Neuroradiol       Date:  2012-03-15       Impact factor: 3.825

2.  Treatment and follow-up of 22 unruptured wide-necked intracranial aneurysms of the internal carotid artery with Onyx HD 500.

Authors:  Werner Weber; Ralf Siekmann; Bernhard Kis; Dietmar Kuehne
Journal:  AJNR Am J Neuroradiol       Date:  2005-09       Impact factor: 3.825

3.  Horizontal stent placement plus coiling in a broad-based basilar-tip aneurysm: an alternative to the Y-stent technique.

Authors:  Isabel Wanke; Elke Gizewski; Michael Forsting
Journal:  Neuroradiology       Date:  2006-08-31       Impact factor: 2.804

4.  Late angiographic and clinical follow-up results of 100 consecutive aneurysms treated with Onyx reconstruction: largest single-center experience.

Authors:  H Saruhan Cekirge; Isil Saatci; M Halil Ozturk; Barbaros Cil; Anil Arat; Michel Mawad; Fikret Ergungor; Deniz Belen; Uygur Er; Sami Turk; Murat Bavbek; Zeki Sekerci; Ethem Beskonakli; Osman E Ozcan; Tuncalp Ozgen
Journal:  Neuroradiology       Date:  2006-01-04       Impact factor: 2.804

Review 5.  Intracranial aneurysms.

Authors:  Roberto Gasparotti; Roberto Liserre
Journal:  Eur Radiol       Date:  2005-01-28       Impact factor: 5.315

6.  Treatment options for wide-necked intracranial aneurysms using a self-expandable hydrophilic coil and a self-expandable stent combination.

Authors:  Jonathan L Brisman; Joon K Song; Yasunari Niimi; Alejandro Berenstein
Journal:  AJNR Am J Neuroradiol       Date:  2005-05       Impact factor: 3.825

7.  Delayed migration of a self-expanding intracranial microstent.

Authors:  M E Kelly; R D Turner; S I Moskowitz; V Gonugunta; M S Hussain; D Fiorella
Journal:  AJNR Am J Neuroradiol       Date:  2008-08-21       Impact factor: 3.825

Review 8.  Current diagnostic approaches to subarachnoid haemorrhage.

Authors:  Jean Marie U-King-Im; Brendan Koo; Rikin A Trivedi; Nicholas J Higgins; Keng Y Tay; Justin J Cross; Nagui M Antoun; Jonathan H Gillard
Journal:  Eur Radiol       Date:  2005-02-12       Impact factor: 5.315

9.  Effect of endovascular treatment on headache in elderly patients with unruptured intracranial aneurysms.

Authors:  D-Q Gu; C-Z Duan; X-F Li; X-Y He; L-F Lai; S-X Su
Journal:  AJNR Am J Neuroradiol       Date:  2012-12-06       Impact factor: 3.825

10.  A new canine carotid artery bifurcation aneurysm model for the evaluation of neurovascular devices.

Authors:  O Naggara; T E Darsaut; I Salazkin; G Soulez; F Guilbert; D Roy; A Weill; G Gevry; J Raymond
Journal:  AJNR Am J Neuroradiol       Date:  2009-12-17       Impact factor: 3.825

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.