Monitor unit calculation for Monte Carlo treatment planning.

C M Ma1, R A Price, J S Li, L Chen, L Wang, E Fourkal, L Qin, J Yang.   

Abstract

In this work, we investigate a formalism for monitor unit (MU) calculation in Monte Carlo based treatment planning. By relating MU to dose measured under reference calibration conditions (central axis, depth of dose maximum in water, 10 cm x 10 cm field defined at 100 cm source-to-surface distance) our formalism determines the MU required for a treatment plan based on the prescription dose and Monte Carlo calculated dose distribution. Detailed descriptions and formulae are given for various clinical situations including conventional treatments and advanced techniques such as intensity-modulated radiotherapy (IMRT) and modulated electron radiotherapy (MERT). Analysis is made of the effects of source modelling, beam modifier simulation and patient dose calculation accuracy, all of which are important factors for absolute dose calculations using Monte Carlo simulations. We have tested the formalism through phantom measurements and the predicted MU values were consistent with measured values to within 2%. The formalism has been used for MU calculation and plan comparison for advanced treatment techniques such as MERT, extracranial stereotactic IMRT, MRI-based treatment planning and intensity-modulated laser-proton therapy studies. It is also used for absolute dose calculations using Monte Carlo simulations for treatment verification, which has become part of our comprehensive IMRT quality assurance programme.

Entities:  

Mesh:

Year:  2004        PMID: 15152923     DOI: 10.1088/0031-9155/49/9/006

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  4 in total

1.  Monte Carlo calculations for absolute dosimetry to determine machine outputs for proton therapy fields.

Authors:  Harald Paganetti
Journal:  Phys Med Biol       Date:  2006-05-17       Impact factor: 3.609

2.  Monitor unit calculations for external photon and electron beams: Report of the AAPM Therapy Physics Committee Task Group No. 71.

Authors:  John P Gibbons; John A Antolak; David S Followill; M Saiful Huq; Eric E Klein; Kwok L Lam; Jatinder R Palta; Donald M Roback; Mark Reid; Faiz M Khan
Journal:  Med Phys       Date:  2014-03       Impact factor: 4.071

3.  Evaluation of dose prediction errors and optimization convergence errors of deliverable-based head-and-neck IMRT plans computed with a superposition/convolution dose algorithm.

Authors:  I B Mihaylov; J V Siebers
Journal:  Med Phys       Date:  2008-08       Impact factor: 4.071

4.  Independent calculation of dose from a helical TomoTherapy unit.

Authors:  John P Gibbons; Koren Smith; Dennis Cheek; Isaac Rosen
Journal:  J Appl Clin Med Phys       Date:  2009-02-05       Impact factor: 2.102

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.