Literature DB >> 15088170

Multislice CT coronary angiography: evaluation of an automatic vessel detection tool.

M Dewey1, D Schnapauff, M Laule, A Lembcke, A C Borges, W Rutsch, B Hamm, P Rogalla.   

Abstract

PURPOSE: To investigate the potential of a new detection tool for multislice CT (MSCT) coronary angiography with automatic display of curved multiplanar reformations and orthogonal cross-sections.
MATERIALS AND METHODS: Thirty-five patients were consecutively enrolled in a prospective intention-to-diagnose study and examined using a MSCT scanner with 16 x 0.5 mm detector collimation and 400 ms gantry rotation time (Aquilion, Toshiba). A multisegment algorithm using up to four segments was applied for ECG-gated reconstruction. Automatic and manual detection of coronary arteries was conducted using the coronary artery CT protocol of a workstation (Vitrea 2, Version 3.3, Vital Images) to detect significant stenoses (> or = 50 %) in all segments of > or = 1.5 mm in diameter. Each detection tool was used by one reader who was blinded to the results of the other detection method and the results of conventional coronary angiography.
RESULTS: The overall sensitivity, specificity, nondiagnostic rate, and accuracy of the automatic and manual approach were 90 vs. 94 %, 89 vs. 84 %, 6 vs. 6 %, and 89 vs. 88 %, respectively (p = n. s.). The vessel length detected with the automatic and manual approach were highly correlated for the left main/left anterior descending (143 +/- 30 vs. 146 +/- 24 mm, r = 0.923, p < 0.001), left circumflex (94 +/- 35 vs. 93 +/- 33 mm, r = 0.945, p < 0.001), and right coronary artery (145 +/- 36 vs. 144 +/- 37 mm, r = 0.925, p < 0.001). The time required to create reformations along the coronary arteries was significantly shorter with the automatic tool compared to the manual approach (203 +/- 77 vs. 391 +/- 104 sec, p < 0.005). In 90 % of the coronary branches automatic detection required less time than the manual approach.
CONCLUSION: Automatic coronary vessel detection is feasible and reduces the time required to create reformations by a factor of approximately two without deteriorating the diagnostic accuracy.

Entities:  

Mesh:

Year:  2004        PMID: 15088170     DOI: 10.1055/s-2004-812991

Source DB:  PubMed          Journal:  Rofo        ISSN: 1438-9010


  23 in total

1.  Coronary arteries: assessment of image quality and optimal reconstruction window in retrospective ECG-gated multislice CT at 375-ms gantry rotation time.

Authors:  Xavier L Hamoir; Thomas Flohr; Vincent Hamoir; Laurent Labaki; Jean-Yves Tricquet; Alain Duhamel; Jacques Kirsch
Journal:  Eur Radiol       Date:  2004-11-24       Impact factor: 5.315

2.  Evaluation of a semiautomatic software tool for left ventricular function analysis with 16-slice computed tomography.

Authors:  Marc Dewey; Mira Müller; Florian Teige; Bernd Hamm
Journal:  Eur Radiol       Date:  2005-06-17       Impact factor: 5.315

3.  Localizing calcifications in cardiac CT data sets using a new vessel segmentation approach.

Authors:  Stefan Wesarg; M Fawad Khan; Evelyn A Firle
Journal:  J Digit Imaging       Date:  2006-09       Impact factor: 4.056

Review 4.  A practical guide to reading CT coronary angiograms--how to avoid mistakes when assessing for coronary stenoses.

Authors:  John W M Hoe; Kok Hong Toh
Journal:  Int J Cardiovasc Imaging       Date:  2006-12-21       Impact factor: 2.357

5.  Multislice computed tomography: angiographic emulation versus standard assessment for detection of coronary stenoses.

Authors:  Dirk Schnapauff; Hans-Peter Dübel; Jürgen Scholze; Gert Baumann; Bernd Hamm; Marc Dewey
Journal:  Eur Radiol       Date:  2007-01-06       Impact factor: 5.315

6.  CT angiography in highly calcified arteries: 2D manual vs. modified automated 3D approach to identify coronary stenoses.

Authors:  Marco A S Cordeiro; Albert C Lardo; Marcelo S V Brito; Miguel A Rosário Neto; Maria H A Siqueira; José R Parga; Luiz F Avila; José A F Ramires; João A C Lima; Carlos E Rochitte
Journal:  Int J Cardiovasc Imaging       Date:  2006-03-15       Impact factor: 2.357

7.  Facilitating coronary artery evaluation in MDCT using a 3D automatic vessel segmentation tool.

Authors:  M Fawad Khan; Stefan Wesarg; Jessen Gurung; Selami Dogan; Adel Maataoui; Boris Brehmer; Christopher Herzog; Hanns Ackermann; Birgit Assmus; Thomas J Vogl
Journal:  Eur Radiol       Date:  2006-03-10       Impact factor: 5.315

Review 8.  Comparison between the image quality of multisegment and halfscan reconstructions of non-invasive CT coronary angiography.

Authors:  D Schnapauff; F Teige; B Hamm; M Dewey
Journal:  Br J Radiol       Date:  2009-06-08       Impact factor: 3.039

9.  Efficacy of computer aided analysis in detection of significant coronary artery stenosis in cardiac using dual source computed tomography.

Authors:  Anja J Reimann; Ilias Tsiflikas; Harald Brodoefel; Michael Scheuering; Daniel Rinck; Andreas F Kopp; Claus D Claussen; Martin Heuschmid
Journal:  Int J Cardiovasc Imaging       Date:  2008-09-28       Impact factor: 2.357

10.  Coronary CT angiography using 64 detector rows: methods and design of the multi-centre trial CORE-64.

Authors:  Julie M Miller; Marc Dewey; Andrea L Vavere; Carlos E Rochitte; Hiroyuki Niinuma; Armin Arbab-Zadeh; Narinder Paul; John Hoe; Albert de Roos; Kunihiro Yoshioka; Pedro A Lemos; David E Bush; Albert C Lardo; John Texter; Jeffery Brinker; Christopher Cox; Melvin E Clouse; João A C Lima
Journal:  Eur Radiol       Date:  2008-11-08       Impact factor: 5.315

View more