Literature DB >> 15087724

Neurostimulation system used for deep brain stimulation (DBS): MR safety issues and implications of failing to follow safety recommendations.

Ali R Rezai1, Michael Phillips, Kenneth B Baker, Ashwini D Sharan, John Nyenhuis, Jean Tkach, Jaimie Henderson, Frank G Shellock.   

Abstract

The use of magnetic resonance imaging (MRI) in patients with neurostimulation systems used for deep brain stimulation requires the utmost care, and no individual should undergo an MR examination in the absence of empirical evidence that the procedure can be performed safely. The risks of performing MRI in patients with neurostimulators include those associated with heating, magnetic field interactions, induced currents, and the functional disruption of these devices. The exact safety recommendations for the particular neurostimulation system with regard to the pulse generator, leads, electrodes, operational conditions for the device, the positioning of these components, and the MR system conditions must be carefully followed for MRI. As highlighted by 2 recent accidents, the failure to strictly follow safety recommendations (eg, use a 1.5-T MR system with a send/receive head radiofrequency coil only; limit the specific absorption rate to 0.4 W/kg; etc.) may result in serious, temporary, or permanent injury to the patient including the possibility of transient dystonia, paralysis, coma, or even death.

Entities:  

Mesh:

Year:  2004        PMID: 15087724     DOI: 10.1097/01.rli.0000124940.02340.ab

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  38 in total

1.  Analysis of the role of lead resistivity in specific absorption rate for deep brain stimulator leads at 3T MRI.

Authors:  Leonardo M Angelone; Jyrki Ahveninen; John W Belliveau; Giorgio Bonmassar
Journal:  IEEE Trans Med Imaging       Date:  2010-03-22       Impact factor: 10.048

2.  Pallidal neuronal discharge in Huntington's disease: support for selective loss of striatal cells originating the indirect pathway.

Authors:  Philip A Starr; Gail A Kang; Susan Heath; Shoichi Shimamoto; Robert S Turner
Journal:  Exp Neurol       Date:  2008-02-14       Impact factor: 5.330

3.  Radiofrequency-induced heating near fixed orthodontic appliances in high field MRI systems at 3.0 Tesla.

Authors:  Marc Regier; Jörn Kemper; Michael G Kaul; Markus Feddersen; Gerhard Adam; Bärbel Kahl-Nieke; Arndt Klocke
Journal:  J Orofac Orthop       Date:  2009-12-04       Impact factor: 1.938

Review 4.  MRI-related heating near deep brain stimulation electrodes: more data are needed.

Authors:  Akshay A Gupte; Devashish Shrivastava; Maggie A Spaniol; Aviva Abosch
Journal:  Stereotact Funct Neurosurg       Date:  2011-04-13       Impact factor: 1.875

5.  Compatibility of temporary pacemaker myocardial pacing leads with magnetic resonance imaging: an ex vivo tissue study.

Authors:  Alexander Pfeil; Stefanie Drobnik; Reinhard Rzanny; Anas Aboud; Joachim Böttcher; Peter Schmidt; Christian Ortmann; Gita Mall; Khosro Hekmat; Bernhard Brehm; Juergen Reichenbach; Thomas E Mayer; Gunter Wolf; Andreas Hansch
Journal:  Int J Cardiovasc Imaging       Date:  2011-02-24       Impact factor: 2.357

6.  Controlling radiofrequency-induced currents in guidewires using parallel transmit.

Authors:  Maryam Etezadi-Amoli; Pascal Stang; Adam Kerr; John Pauly; Greig Scott
Journal:  Magn Reson Med       Date:  2014-12-17       Impact factor: 4.668

7.  Construction and modeling of a reconfigurable MRI coil for lowering SAR in patients with deep brain stimulation implants.

Authors:  Laleh Golestanirad; Maria Ida Iacono; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Michael D Fox; Todd Herrington; Elfar Adalsteinsson; Cristen LaPierre; Azma Mareyam; Lawrence L Wald
Journal:  Neuroimage       Date:  2016-12-21       Impact factor: 6.556

8.  Feasibility of using linearly polarized rotating birdcage transmitters and close-fitting receive arrays in MRI to reduce SAR in the vicinity of deep brain simulation implants.

Authors:  Laleh Golestanirad; Boris Keil; Leonardo M Angelone; Giorgio Bonmassar; Azma Mareyam; Lawrence L Wald
Journal:  Magn Reson Med       Date:  2016-04-05       Impact factor: 4.668

Review 9.  Post-operative imaging in deep brain stimulation: A controversial issue.

Authors:  Christian Saleh; Georges Dooms; Christophe Berthold; Frank Hertel
Journal:  Neuroradiol J       Date:  2016-03-30

10.  Serial (1)H-MRS of thalamus during deep brain stimulation of bilateral globus pallidus internus for primary generalized dystonia.

Authors:  Mikhail F Chernov; Taku Ochiai; Takaomi Taira; Yuko Ono; Ryoichi Nakamura; Yoshihiro Muragaki; Hiroshi Iseki; Tomokatsu Hori; Kintomo Takakura
Journal:  Neuroradiology       Date:  2008-09-30       Impact factor: 2.804

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.