Noninvasive inductive stent heating: alternative approach to prevent instent restenosis?

Michael G Floren1, Rolf W Günther, Thomas Schmitz-Rode.   

Abstract

RATIONALE AND
OBJECTIVES: To test noninvasive inductive heating of implanted vascular stents as an alternative approach for reduction or prevention of neointimal hyperplasia.
METHODS: Calorimetric pretests were performed to get an orientation on the different parameters of influence for inductive heating of stents. The field strength was set to a maximum of 90 kA/m within a frequency range from 80 kHz to 320 kHz. The electromagnetic field was emitted by a custom-made water-cooled copper winding antenna. A flow model for stent heating was set up to assess the increase in temperature of an expanded 316L stainless steel stent with typical coronary stent dimensions of 3.5 mm diameter and 14.5 mm in length, and in a second setup with 4.5 mm diameter and 13 mm in length, respectively. The stent was located in a bioartificial artery, simulated by a fibrinogen matrix with a defined number of vital cells. The system was exposed to a pulsating perfusion and to an electromagnetic field of 200 kHz over a period of 20 minutes and in a second setup to an electromagnetic field of 300 kHz and increasing intensity up to maximum power-output. Afterward, the artificial vessel was sliced and examined by fluorescence microscopy to evaluate the number and location of damaged cells.
RESULTS: The calorimetric tests show an exponential correlation of energy uptake in the stent with an increase in frequency and a constant generator output. At a frequency of 80 kHz, the power uptake accounts for 0.1 W (250 kHz 1.0 W; 320 kHz 1.9 W, respectively). The flow tests confirmed feasibility to elevate the stent temperature from 37 degrees C body temperature to 44 degrees C at 200 kHz within 55 seconds. The temperature increase of the fluid passing the heated vessel region was only marginal (maximum of 0.5 degrees C). Cell necrosis after 20 minutes of treatment was not observed. In a second set-up with 4.5 mm stent diameter, a frequency of 300 kHz and with maximum power output, the stent temperature was increased to 80 degrees C and there was extensive necrosis area around the stent. Treatment time and stent temperature were optimized in further tests.
CONCLUSION: Selective noninvasive energy transfer to coronary stainless steel stents by inductive heating is possible within a wide range of power. By thermal conduction, vital cells close to the stent struts can be affected. The frequency of 200 kHz turned out to be favorable. There is still room for further optimization of energy dosage with regard to material and stent design, to induce controlled cell death. The method has potential to serve as an alternative approach for prevention of instent restenosis.

Entities:  

Mesh:

Year:  2004        PMID: 15087720     DOI: 10.1097/01.rli.0000117881.42422.72

Source DB:  PubMed          Journal:  Invest Radiol        ISSN: 0020-9996            Impact factor:   6.016


  5 in total

1.  Collagen-cellulose composite thin films that mimic soft-tissue and allow stem-cell orientation.

Authors:  Terry W J Steele; Charlotte L Huang; Evelyne Nguyen; Udi Sarig; Saranya Kumar; Effendi Widjaja; Joachim S C Loo; Marcelle Machluf; Freddy Boey; Zlata Vukadinovic; Andreas Hilfiker; Subbu S Venkatraman
Journal:  J Mater Sci Mater Med       Date:  2013-05-14       Impact factor: 3.896

2.  A Novel Shape Memory Plate Osteosynthesis for Noninvasive Modulation of Fixation Stiffness in a Rabbit Tibia Osteotomy Model.

Authors:  Christian W Müller; Ronny Pfeifer; Karen Meier; Sebastian Decker; Janin Reifenrath; Thomas Gösling; Volker Wesling; Christian Krettek; Christof Hurschler; Manuel Krämer
Journal:  Biomed Res Int       Date:  2015-06-08       Impact factor: 3.411

3.  In vitro study on the feasibility of magnetic stent hyperthermia for the treatment of cardiovascular restenosis.

Authors:  Li Li; Rui Wang; Huan-Huan Shi; LE Xie; Jing-Ding-Sha Li; Wei-Chao Kong; Jin-Tian Tang; DA-Nian Ke; Ling-Yun Zhao
Journal:  Exp Ther Med       Date:  2013-06-21       Impact factor: 2.447

Review 4.  Health Care Monitoring and Treatment for Coronary Artery Diseases: Challenges and Issues.

Authors:  Mokhalad Alghrairi; Nasri Sulaiman; Saad Mutashar
Journal:  Sensors (Basel)       Date:  2020-08-01       Impact factor: 3.576

5.  Nanomagnetic Actuation of Hybrid Stents for Hyperthermia Treatment of Hollow Organ Tumors.

Authors:  Benedikt Mues; Benedict Bauer; Anjali A Roeth; Jeanette Ortega; Eva Miriam Buhl; Patricia Radon; Frank Wiekhorst; Thomas Gries; Thomas Schmitz-Rode; Ioana Slabu
Journal:  Nanomaterials (Basel)       Date:  2021-03-02       Impact factor: 5.076

  5 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.