Evaluation of immunochemiluminometric assays for the measurement of insulin and C-peptide using the ADVIA Centaur.

Hanane El Kenz1, Pierre Bergmann.   

Abstract

We evaluated new immunochemiluminometric assays (ICMAs) for insulin and C-peptide (ADVIA Centaur Insulin &amp; C-Peptide-Serum assays). Both ADVIA Centaur assays are two-site sandwich immunoassays using direct chemiluminescent technology. Precision was investigated using serum pools at three levels of the two analytes, measured in duplicate for 10 days. Total Coefficient of Variations (CVs) were 5, 7 and 4% for insulin and 9, 6 and 10% for C-peptide, with intra-assay precisions of 5, 4 and 5% and 5, 3 and 3%, respectively. The minimum detectable concentrations were 0.5 mU/L and < 0.1 microg/L for insulin and for C-peptide, respectively. Day-to-day reproducibility of single measurements was 5.4, 7.1 and 4.3% for pools with an insulin concentration of 0.6 mU/L, 2.0 mU/L and 4.0 mU/L; it was 4.4, 6.6 and 5.3% for pools with a C-peptide concentration of 0.2, 0.3 and 1.0 microg/L. The functional sensitivity did not differ from 3 SD Minimal Detectable Concentration (MDC) (0.5 mU/L for insulin and < 0.1 microg/L for C-peptide). The linearity was good in the range of 0.6-20 mU/L for insulin and 0.3-9 microg/L for C-peptide. The comparison with the RIA used in our laboratory was analyzed by Passing-Bablok and Bland-Altman plots and revealed a proportional bias of approximately 20% (slope: 1.20; CI: 1.14 to 1.26) for C-peptide and a systematic bias of -1.6 mU/L (slope: 0.94; CI: 2.7 to -0.5) for insulin which should not have any clinical consequence in the interpretation of results. Finally, we tested the influence of hemolysis on insulin in serum and plasma and found the same negative effect for both samples when more than 2% of red cells were hemolyzed, and this effect increased with the lag time before freezing. In conclusion, both assays were satisfactorily correlated with the routine RIA test used in our laboratory. The major problem was the sensitivity to hemolysis which is common to all insulin immunometric assays.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15074471

Source DB:  PubMed          Journal:  Clin Lab        ISSN: 1433-6510            Impact factor:   1.138


  4 in total

1.  Individualized correction of insulin measurement in hemolyzed serum samples.

Authors:  Zhi-Qi Wu; Ju Lu; Huanhuan Chen; Wensen Chen; Hua-Guo Xu
Journal:  Immunol Res       Date:  2017-06       Impact factor: 2.829

2.  Menopausal Vasomotor Symptoms and Risk of Incident Cardiovascular Disease Events in SWAN.

Authors:  Rebecca C Thurston; Helen E Aslanidou Vlachos; Carol A Derby; Elizabeth A Jackson; Maria Mori Brooks; Karen A Matthews; Sioban Harlow; Hadine Joffe; Samar R El Khoudary
Journal:  J Am Heart Assoc       Date:  2021-01-20       Impact factor: 5.501

3.  Cardiovascular Disease Risk Factor Burden During the Menopause Transition and Late Midlife Subclinical Vascular Disease: Does Race/Ethnicity Matter?

Authors:  Emma Barinas-Mitchell; Chunzhe Duan; Maria Brooks; Samar R El Khoudary; Rebecca C Thurston; Karen A Matthews; Elizabeth A Jackson; Tené T Lewis; Carol A Derby
Journal:  J Am Heart Assoc       Date:  2020-02-17       Impact factor: 5.501

4.  Interpersonal Trauma and Risk of Incident Cardiovascular Disease Events Among Women.

Authors:  Rebecca C Thurston; Yuefang Chang; Karen A Matthews; Sioban Harlow; Samar R El Khoudary; Imke Janssen; Carol Derby
Journal:  J Am Heart Assoc       Date:  2022-03-24       Impact factor: 5.501

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.