Variability of maximal aortic aneurysm diameter measurements on CT scan: significance and methods to minimize.

Neal S Cayne1, Frank J Veith, Evan C Lipsitz, Takao Ohki, Manish Mehta, Nick Gargiulo, William D Suggs, Alla Rozenblit, Zina Ricci, Carlos H Timaran.   

Abstract

OBJECTIVES: We noted substantial differences when measuring repeatedly the same abdominal aortic aneurysm (AAA) on the same computed tomography (CT) scan. This study quantitated this variability, and methods to minimize it were developed.
METHODS: The CT maximal diameter of 25 AAAs was measured by eight experienced observers, including six vascular surgeons and two radiologists, using two methods: an unstandardized protocol, and a standardized protocol using fine calipers to carefully measure the largest diameter perpendicular to the estimated aneurysm centerline, from outer aneurysm wall to outer wall. The average measurement difference between observers was calculated for each method. The average difference between each observer's measurement and the official radiology report value was also calculated. Agreement between the two measurement methods was assessed with Bland-Altman plots.
RESULTS: The difference in maximal diameter measurements between each observer averaged 4.0 +/- 5.1 mm (range, 0.0-35.0 mm) with the unstandardized method. The mean measurement difference with the standardized protocol was significantly lower, and averaged 2.8 +/- 4.4 mm (range, 0.0-26.0 mm; P<.05). Measurements taken from the official radiology report differed from each of the observer's standardized measurement by an average of 5.0 +/- 6.3 mm (range, 0.0-28.0 mm). This difference was similar for both the unstandardized and standardized methods. Bland-Altman plots confirmed the wide variation of the maximal diameter measurements when the unstandardized method was compared with the standardized method (95% confidence interval, -9-9 mm).
CONCLUSIONS: Routine CT maximal diameter measurement of AAAs can have substantial interobserver variability. Standardized measurement protocols can decrease, but not eliminate, this measurement variability. Thus apparent size changes based on CT measurements may represent measurement artifact rather than actual aneurysm growth or shrinkage, particularly when a standardized system is not used.

Entities:  

Mesh:

Year:  2004        PMID: 15071447     DOI: 10.1016/j.jvs.2003.11.042

Source DB:  PubMed          Journal:  J Vasc Surg        ISSN: 0741-5214            Impact factor:   4.268


  8 in total

1.  Sizing for endovascular aneurysm repair: clinical evaluation of a new automated three-dimensional software.

Authors:  Adrien Kaladji; Antoine Lucas; Gaëlle Kervio; Pascal Haigron; Alain Cardon
Journal:  Ann Vasc Surg       Date:  2010-10       Impact factor: 1.466

2.  Prevalence and significance of incidentally noted dilation of the ascending aorta on routine chest computed tomography in older patients.

Authors:  Nancy Benedetti; Michael D Hope
Journal:  J Comput Assist Tomogr       Date:  2015 Jan-Feb       Impact factor: 1.826

3.  Usefulness of new imaging methods for assessment of type B aortic dissection.

Authors:  Rachel E Clough; Vassilios E Zymvragoudakis; Lukla Biasi; Peter R Taylor
Journal:  Ann Cardiothorac Surg       Date:  2014-05

4.  Maximum Diameter of Native Abdominal Aortic Aneurysm Measured by Angio-Computed Tomography: Reproducibility and Lack of Consensus Impacts on Clinical Decisions.

Authors:  Caroline E Mora; Claude D Marcus; Coralie M Barbe; Fiona B Ecarnot; Anne L Long
Journal:  Aorta (Stamford)       Date:  2015-04-01

5.  A Deep Learning Approach to Visualise Aortic Aneurysm Morphology without the Use of Intravenous Contrast Agents.

Authors:  Anirudh Chandrashekar; Ashok Handa; Pierfrancesco Lapolla; Natesh Shivakumar; Raman Uberoi; Vicente Grau; Regent Lee
Journal:  Ann Surg       Date:  2021-03-04       Impact factor: 12.969

6.  Comparison of colour duplex ultrasound with computed tomography to measure the maximum abdominal aortic aneurysmal diameter.

Authors:  C Gray; P Goodman; S A Badger; M K O'Malley; M K O'Donohoe; C O McDonnell
Journal:  Int J Vasc Med       Date:  2014-11-23

7.  Evaluation of angiography as the sole imaging study for the proximal aortic neck prior to EVAR.

Authors:  Stephen A Badger; Nityanda Arya; William Loan; Chee V Soong
Journal:  Ulster Med J       Date:  2009-09

8.  Vascular Deformation Mapping of Abdominal Aortic Aneurysm.

Authors:  Drew J Braet; Jonathan Eliason; Yunus Ahmed; Pieter A J van Bakel; Jiayang Zhong; Zhangxing Bian; Carlos Alberto Figueroa; Nicholas S Burris
Journal:  Tomography       Date:  2021-05-13
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.