Literature DB >> 15071001

Evaluation of the automated phoenix system for potential routine use in the clinical microbiology laboratory.

J-L Donay1, D Mathieu, P Fernandes, C Prégermain, P Bruel, A Wargnier, I Casin, F X Weill, P H Lagrange, J L Herrmann.   

Abstract

A comparative study was designed to evaluate the identification (ID) and antimicrobial susceptibility testing (AST) performances of the BD Phoenix Automated Microbiology System (Becton Dickinson Diagnostic Systems [BD], Pont de Claix, France). A total of 305 single clinical isolates were collected, and comparisons were made with routine manual methods in use in our microbiology laboratories. The percentages of correct IDs were 93.3, 89.4, 91.8, and 85.7% for enterobacteria, nonfermenting gram-negative bacilli, staphylococci, and streptococci-enterococci, respectively. The median ID time was 3 h, and the median time for AST was 10 h 30 min. AST results showed variable percentages of errors for the different antibiotics. None of the enterobacteria and 0.3% of Pseudomonas aeruginosa isolates showed a very major error (VME). Only one strain of Staphylococcus aureus showed a VME with oxacillin. We demonstrate here the efficiency of the Phoenix system, which can be used for the majority of strains encountered in a university-based laboratory, for ID and AST.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15071001      PMCID: PMC387561          DOI: 10.1128/JCM.42.4.1542-1546.2004

Source DB:  PubMed          Journal:  J Clin Microbiol        ISSN: 0095-1137            Impact factor:   5.948


  42 in total

1.  Evaluation of an automated system for identification of Enterobacteriaceae and nonfermenting bacilli.

Authors:  M F Jossart; R J Courcol
Journal:  Eur J Clin Microbiol Infect Dis       Date:  1999-12       Impact factor: 3.267

2.  [Antibiogram Committee of the French Microbiology Society. Report 2000-2001].

Authors:  C J Soussy; G Carret; J D Cavallo; H Chardon; C Chidiac; P Choutet; P Courvalin; H Dabernat; H Drugeon; L Dubreuil; F Goldstein; V Jarlier; R Leclercq; M H Nicolas-Chanoine; A Philippon; C Quentin; B Rouveix; J Sirot
Journal:  Pathol Biol (Paris)       Date:  2000-11

Review 3.  Multiple antibiotic resistance and efflux.

Authors:  H Nikaido
Journal:  Curr Opin Microbiol       Date:  1998-10       Impact factor: 7.934

4.  Evaluation of VITEK 2 rapid identification and susceptibility testing system against gram-negative clinical isolates.

Authors:  T K Ling; P C Tam; Z K Liu; A F Cheng
Journal:  J Clin Microbiol       Date:  2001-08       Impact factor: 5.948

5.  Potential impact of the VITEK 2 system and the Advanced Expert System on the clinical laboratory of a university-based hospital.

Authors:  C C Sanders; M Peyret; E S Moland; S J Cavalieri; C Shubert; K S Thomson; J M Boeufgras; W E Sanders
Journal:  J Clin Microbiol       Date:  2001-07       Impact factor: 5.948

Review 6.  Efflux-mediated resistance to fluoroquinolones in gram-negative bacteria.

Authors:  K Poole
Journal:  Antimicrob Agents Chemother       Date:  2000-09       Impact factor: 5.191

7.  Evaluation of a new system, VITEK 2, for identification and antimicrobial susceptibility testing of enterococci.

Authors:  F Garcia-Garrote; E Cercenado; E Bouza
Journal:  J Clin Microbiol       Date:  2000-06       Impact factor: 5.948

8.  Outbreak of Klebsiella pneumoniae producing transferable AmpC-type beta-lactamase (ACC-1) originating from Hafnia alvei.

Authors:  D Nadjar; M Rouveau; C Verdet; L Donay; J Herrmann; P H Lagrange; A Philippon; G Arlet
Journal:  FEMS Microbiol Lett       Date:  2000-06-01       Impact factor: 2.742

9.  Accuracy of the VITEK 2 system to detect glycopeptide resistance in enterococci.

Authors:  N van Den Braak; W Goessens; A van Belkum; H A Verbrugh; H P Endtz
Journal:  J Clin Microbiol       Date:  2001-01       Impact factor: 5.948

10.  Comparison of the Vitek Gram-Positive Susceptibility 106 card and the MRSA-screen latex agglutination test for determining oxacillin resistance in clinical bloodstream isolates of Staphylococcus aureus.

Authors:  T Yamazumi; S A Marshall; W W Wilke; D J Diekema; M A Pfaller; R N Jones
Journal:  J Clin Microbiol       Date:  2001-01       Impact factor: 5.948

View more
  36 in total

1.  Comparison of Bruker Biotyper matrix-assisted laser desorption ionization-time of flight mass spectrometer to BD Phoenix automated microbiology system for identification of gram-negative bacilli.

Authors:  Ryan T Saffert; Scott A Cunningham; Sherry M Ihde; Kristine E Monson Jobe; Jayawant Mandrekar; Robin Patel
Journal:  J Clin Microbiol       Date:  2011-01-05       Impact factor: 5.948

2.  Evaluation of the BD Phoenix automated microbiology system for identification and antimicrobial susceptibility testing of staphylococci and enterococci.

Authors:  Karen C Carroll; Anita P Borek; Chad Burger; Brian Glanz; Hasan Bhally; Susan Henciak; Diane C Flayhart
Journal:  J Clin Microbiol       Date:  2006-06       Impact factor: 5.948

Review 3.  Manual and automated instrumentation for identification of Enterobacteriaceae and other aerobic gram-negative bacilli.

Authors:  Caroline M O'hara
Journal:  Clin Microbiol Rev       Date:  2005-01       Impact factor: 26.132

4.  Use of the MicroSeq 500 16S rRNA gene-based sequencing for identification of bacterial isolates that commercial automated systems failed to identify correctly.

Authors:  Carla Fontana; Marco Favaro; Marco Pelliccioni; Enrico Salvatore Pistoia; Cartesio Favalli
Journal:  J Clin Microbiol       Date:  2005-02       Impact factor: 5.948

5.  Analysis of the comparative workflow and performance characteristics of the VITEK 2 and Phoenix systems.

Authors:  U Eigner; A Schmid; U Wild; D Bertsch; A-M Fahr
Journal:  J Clin Microbiol       Date:  2005-08       Impact factor: 5.948

6.  Two-center collaborative evaluation of performance of the BD phoenix automated microbiology system for identification and antimicrobial susceptibility testing of gram-negative bacteria.

Authors:  Maria Grazia Menozzi; Ulrich Eigner; Silvia Covan; Sabina Rossi; Pietro Somenzi; Giuseppe Dettori; Carlo Chezzi; Anne-Marie Fahr
Journal:  J Clin Microbiol       Date:  2006-09-27       Impact factor: 5.948

7.  Accuracies of beta-lactam susceptibility test results for Pseudomonas aeruginosa with four automated systems (BD Phoenix, MicroScan WalkAway, Vitek, and Vitek 2).

Authors:  Stefan Juretschko; Vincent J Labombardi; Stephen A Lerner; Paul C Schreckenberger
Journal:  J Clin Microbiol       Date:  2007-01-17       Impact factor: 5.948

8.  Isolation of Pseudomonas aeruginosa from open ocean and comparison with freshwater, clinical, and animal isolates.

Authors:  Nurul Huda Khan; Yoshikazu Ishii; Noriko Kimata-Kino; Hidetake Esaki; Tomohiko Nishino; Masahiko Nishimura; Kazuhiro Kogure
Journal:  Microb Ecol       Date:  2007-02       Impact factor: 4.552

9.  Evaluation of the inoculation procedure using a 0.25 McFarland standard for the BD Phoenix automated microbiology system.

Authors:  J-L Donay; P Fernandes; P H Lagrange; J-L Herrmann
Journal:  J Clin Microbiol       Date:  2007-10-03       Impact factor: 5.948

10.  Multi-center evaluation of the VITEK® MS system for mass spectrometric identification of non-Enterobacteriaceae Gram-negative bacilli.

Authors:  R Manji; M Bythrow; J A Branda; C-A D Burnham; M J Ferraro; O B Garner; R Jennemann; M A Lewinski; A B Mochon; G W Procop; S S Richter; J A Rychert; L Sercia; L F Westblade; C C Ginocchio
Journal:  Eur J Clin Microbiol Infect Dis       Date:  2013-09-10       Impact factor: 3.267

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.