Literature DB >> 15066741

A between-patient disinfection method to control water line contamination and biofilm inside dental units.

L Montebugnoli1, S Chersoni, C Prati, G Dolci.   

Abstract

The aim of the present study was to evaluate the efficacy of a between-patient disinfection procedures to maintain low bacterial counts in dental unit water line (DUWL) effluents, and control dental water line biofilms. Six dental units already in use, that had never been cleaned, were monitored for three weeks. During the first week only baseline contamination levels were assessed with no treatment of the system. In the second week lines were flushed with water for 30 s before treating each patient. During the third week, a disinfection procedure with 0.26% peracetic acid, followed by a water flush, was implemented before treating each patient. DUWL samples were collected both at the beginning and at the end of 216 dental procedures (72 during each period), plated on R2A agar and incubated at room temperature for seven days to obtain total bacterial counts in colony forming units per millilitre. To assess biofilm control, nine dental units (five never used and four old dental units with established biofilm) were used for 30 days in routine dental practice undergoing five between-patient DUWL disinfecting cycles every day. Water line samples were removed at baseline and at the end of the study and examined by scanning electron microscopy to determine the presence or absence of biofilms. A significant difference (P < 0.01) in mean DUWL bacterial counts was found between the three sets of observations. Biofilms were not present in any of the new dental units and a demonstrable reduction in the biofilms from the four dental units with previous presence of established biofilms was observed at the end of the study. In this study, a between-patient disinfection procedure consisting of flushing DUWL with peracetic acid with use of water was efficacious in the control of both microbial contamination of dental treatment water and dental water line biofilms.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 15066741     DOI: 10.1016/j.jhin.2004.01.015

Source DB:  PubMed          Journal:  J Hosp Infect        ISSN: 0195-6701            Impact factor:   3.926


  9 in total

1.  Comparison of the microbial load of incoming and distal outlet waters from dental unit water systems in Istanbul.

Authors:  Irfan Türetgen; Duygu Göksay; Aysin Cotuk
Journal:  Environ Monit Assess       Date:  2008-10-09       Impact factor: 2.513

Review 2.  Is peracetic acid suitable for the cleaning step of reprocessing flexible endoscopes?

Authors:  Günter Kampf; Patricia M Fliss; Heike Martiny
Journal:  World J Gastrointest Endosc       Date:  2014-09-16

3.  In vitro comparison of peracetic acid and autoclave sterilization in the corrosion of orthodontic pliers: a pilot study.

Authors:  Livia Kelly Ferraz Nunes; Érika Lopes Carneiro; Nathalia Barbosa Palomares; Lilian Senna; Helio Sampaio-Filho; José Augusto Mendes Miguel
Journal:  Dental Press J Orthod       Date:  2021-10-25

4.  Effect of biocides on biofilm bacteria from dental unit water lines.

Authors:  I Liaqat; A N Sabri
Journal:  Curr Microbiol       Date:  2008-03-06       Impact factor: 2.188

5.  Evaluation of bacterial contamination of dental unit waterlines and use of a newly designed measurement device to assess retraction of a dental chair unit.

Authors:  Xue-Yue Ji; Chun-Nan Fei; Ying Zhang; Wei Zhang; Jun Liu; Jie Dong
Journal:  Int Dent J       Date:  2016-03-21       Impact factor: 2.607

6.  Impact of a risk management plan on Legionella contamination of dental unit water.

Authors:  Erica Leoni; Laura Dallolio; Francesca Stagni; Tiziana Sanna; Giovanni D'Alessandro; Gabriela Piana
Journal:  Int J Environ Res Public Health       Date:  2015-02-23       Impact factor: 3.390

7.  Plasma Sterilization Effectively Reduces Bacterial Contamination in Dental Unit Waterlines.

Authors:  Sarocha Noopan; Phattranit Unchui; Supitcha Techotinnakorn; Ruchanee Salingcarnboriboon Ampornaramveth
Journal:  Int J Dent       Date:  2019-07-30

8.  Marine-Source Quorum Quenching Enzyme YtnP to Improve Hygiene Quality in Dental Units.

Authors:  Xiaohui Sun; Philip Hill; Jia Liu; Jing Qian; Yuting Ma; Shufeng Zhou
Journal:  Mar Drugs       Date:  2021-04-16       Impact factor: 5.118

9.  Effect of different disinfection protocols on microbial and biofilm contamination of dental unit waterlines in community dental practices.

Authors:  Laura Dallolio; Amalia Scuderi; Maria S Rini; Sabrina Valente; Patrizia Farruggia; Maria A Bucci Sabattini; Gianandrea Pasquinelli; Anna Acacci; Greta Roncarati; Erica Leoni
Journal:  Int J Environ Res Public Health       Date:  2014-02-18       Impact factor: 3.390

  9 in total