Abstract

The Allient is a dialysis system that combines various technologies to allow dialysis to be performed at sites outside of dialysis units (intensive care unit [ICU] or home) with ease and safety. A sorbent column regenerates dialysate, removing toxins and providing ultrapure dialysate from only 6 liters of tap water. The use of the sorbent column eliminates the need for costly and complex water purification systems. The Pulsar Blood Movement System provides blood flow at constant negative or positive pressure through single-lumen or dual-lumen accesses, maximizing blood flow rate while eliminating bothersome pressure alarms. Ultrasonic flow monitors control the operation of the pump and ensure adequate blood flow during each dialysis treatment. A completely disposable blood tubing and dialysate circuit eliminates the need for sterilization of the machine. The Allient should make dialysis in the ICU or home setting much more practical, reducing training requirements and increasing safety.

Entities:  

Mesh:

Year:  2004        PMID: 15043624     DOI: 10.1111/j.0894-0959.2004.17219.x

Source DB:  PubMed          Journal:  Semin Dial        ISSN: 0894-0959            Impact factor:   3.455


  6 in total

Review 1.  Renal replacement therapy review: past, present and future.

Authors:  Geoffrey M Fleming
Journal:  Organogenesis       Date:  2011-01-01       Impact factor: 2.500

2.  Removal of Uremic Solutes from Dialysate by Activated Carbon.

Authors:  Seolhyun Lee; Tammy L Sirich; Ignacio J Blanco; Natalie S Plummer; Timothy W Meyer
Journal:  Clin J Am Soc Nephrol       Date:  2022-07-14       Impact factor: 10.614

Review 3.  Water quality in conventional and home haemodialysis.

Authors:  Matthew J Damasiewicz; Kevan R Polkinghorne; Peter G Kerr
Journal:  Nat Rev Nephrol       Date:  2012-10-23       Impact factor: 28.314

4.  Technical breakthroughs in the wearable artificial kidney (WAK).

Authors:  Victor Gura; Alexandra S Macy; Masoud Beizai; Carlos Ezon; Thomas A Golper
Journal:  Clin J Am Soc Nephrol       Date:  2009-08-20       Impact factor: 8.237

5.  Renal replacement therapy in austere environments.

Authors:  Christina M Yuan; Robert M Perkins
Journal:  Int J Nephrol       Date:  2011-04-26

6.  Inflammatory Response to Sorbent Hemodialysis.

Authors:  Laura Rosales; Stephan Thijssen; Anja Kruse; Murat Hairy Sipahioglu; Padam Hirachan; Jochen G Raimann; Viktoriya Kuntsevich; Mary Carter; Nathan W Levin; Peter Kotanko
Journal:  ASAIO J       Date:  2015 Jul-Aug       Impact factor: 2.872

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.