[Method of calculating side scatter factor for diagnostic X-rays].

Hideki Kato1.   

Abstract

When computing the amount of leakage from a diagnostic X-ray room, the amount of scattered radiation released from the irradiated body in the lateral direction must be evaluated correctly. The side-scatter factor varies owing to change in the area and/or shape of the field, but the amount of variation is not always linearly proportional to the area of the field. Because the X-ray spectrum depends on the total filtration of the X-ray tube even if the irradiation geometry, X-ray equipment, and tube potential are the same, the side-scatter factor, too, is subject to change. In this paper, we propose a new method of calculation that uses the differential side-scatter factor computed by means of a Monte Carlo simulation, for obtaining the side-scatter factor of diagnostic X-rays. This method, which can calculate the side-scatter factor under any conditions of irradiation, is useful in evaluating the side-scatter factor of diagnostic X-rays.

Entities:  

Mesh:

Year:  2004        PMID: 15041914     DOI: 10.6009/jjrt.kj00000922265

Source DB:  PubMed          Journal:  Nihon Hoshasen Gijutsu Gakkai Zasshi        ISSN: 0369-4305


  1 in total

1.  Investigation of scatter fractions for estimating leakage dose in medical X-ray imaging facilities.

Authors:  Kimiya Noto; Kichirou Koshida; Hiroji Iida; Tomoyuki Yamamoto; Ikuo Kobayashi; Chikako Kawabata
Journal:  Radiol Phys Technol       Date:  2009-04-08