Endothelial cell integrity after phacoemulsification with 2 different handpieces.

Luis G Vargas1, Mike P Holzer, Kerry D Solomon, Helga P Sandoval, Gerd U Auffarth, David J Apple.   

Abstract

PURPOSE: To investigate corneal endothelial cell loss in fresh phakic human autopsy cadaver eyes after phacoemulsification (PE) with a conventional handpiece or a new oscillatory handpiece.
SETTING: Department of Ophthalmology, Storm Eye Institute, Medical University of South Carolina, Charleston, South Carolina, USA.
METHODS: Twenty fresh phakic human autopsy globes were randomized to PE with a conventional handpiece (n = 10) or an oscillatory handpiece (n = 10). The main parameters evaluated were age, nucleus density, amount of ophthalmic viscosurgical device (OVD) used, and mean PE time and ultrasound (US) power. After surgery, the corneas were excised and the endothelia were vitally stained with trypan blue and alizarin red. Light microscopy was used to assess the endothelial cell loss.
RESULTS: The mean age, density of the nucleus, OVD used, and US power were similar between the groups. The mean PE time was less with the NeoSoniX handpiece (0.71 minutes +/- 0.39 [SD]) than with the conventional handpiece (1.05 +/- 0.53 minutes). The mean number of dead endothelial cells was significantly lower in the oscillatory handpiece group (31.3 +/- 24.2 cells/mm(2)) than in the conventional handpiece group (60.4 +/- 46.8 cells/mm(2)) (P<.001, Mann-Whitney rank sum test).
CONCLUSIONS: Less corneal endothelial cell loss occurred after PE with an oscillatory handpiece than with a conventional handpiece. Further randomized clinical trials are recommended to validate this study.

Entities:  

Mesh:

Year:  2004        PMID: 15030845     DOI: 10.1016/S0886-3350(03)00620-5

Source DB:  PubMed          Journal:  J Cataract Refract Surg        ISSN: 0886-3350            Impact factor:   3.351


  6 in total

1.  Ozil® versus conventional ultrasound phacoemulsification: a randomized comparative study.

Authors:  Afef Maalej; Asma Khallouli; Cheima Wathek; Riadh Rannen
Journal:  Int Ophthalmol       Date:  2014-08-26       Impact factor: 2.031

2.  Corneal endothelial morphology and function after torsional and longitudinal ultrasound mode phacoemulsification.

Authors:  László Jr Módis; Eszter Szalai; Zsuzsa Flaskó; Gábor Németh
Journal:  Rom J Ophthalmol       Date:  2016 Apr-Jun

3.  New ex vivo model of corneal endothelial phacoemulsification injury and rescue therapy with mesenchymal stromal cell secretome.

Authors:  Majid Rouhbakhshzaeri; Behnam Rabiee; Nathalie Azar; Elham Ghahari; Ilham Putra; Medi Eslani; Ali R Djalilian
Journal:  J Cataract Refract Surg       Date:  2018-12-07       Impact factor: 3.351

4.  The effects of blunt trauma and cataract surgery on corneal endothelial cell density.

Authors:  Baris Yeniad; Isik Corum; Cahit Ozgun
Journal:  Middle East Afr J Ophthalmol       Date:  2010-10

5.  Comparative analysis of corneal morphological changes after transversal and torsional phacoemulsification through 2.2 mm corneal incision.

Authors:  Ahmed Assaf; Maged Maher Roshdy
Journal:  Clin Ophthalmol       Date:  2013-01-08

6.  Comparison of effective phacoemulsification time and corneal endothelial cell loss using three different ultrasound frequencies: A randomized controlled trial.

Authors:  Taru Dewan; Praveen K Malik; Preeti Tomar
Journal:  Indian J Ophthalmol       Date:  2022-04       Impact factor: 2.969

  6 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.