Validation of the Pharma-Smart PS-2000 public use blood pressure monitor.

Bruce S Alpert1.   

Abstract

OBJECTIVE: To test the Pharma-Smart Model PS-2000 public use blood pressure monitor for compliance with the Association for the Advancement of Medical Instrumentation's Standard (AAMI) and to a modified British Hypertension Society (BHS) Protocol.
METHODS: Subjects tested ranged in age from 18-74, with the average age of 40. Arm circumference ranged from 22-38 cm. Resting systolic blood pressure (BP) ranged from 91-252 mmHg, and resting diastolic pressure ranged from 57-160 mmHg. There were 44 males, and 41 females. For each subject the readings obtained by the PS-2000 were compared with auscultatory readings obtained by two clinicians, blinded to the results of each other and the device. The manual reference measurements were alternated with the readings obtained by the device.
RESULTS: The average differences between the reference readings (average of the two clinicians) using the AAMI analysis and the automated readings were 0.07+/-7.0 mmHg (Mean+/-SD) for systolic BP, and -0.3+/-6.6 mmHg for diastolic BP.
CONCLUSIONS: The device met the accuracy requirements of the AAMI standard. In addition, when the data were analyzed to assess the compliance with the current British Hypertension Society Protocol (BHS), the device earned the highest rating of 'A' for both systolic and diastolic pressure. We believe that the Pharma-Smart PS-2000 will provide valid readings when placed in non-medical public use sites.

Entities:  

Mesh:

Year:  2004        PMID: 15021074     DOI: 10.1097/00126097-200402000-00005

Source DB:  PubMed          Journal:  Blood Press Monit        ISSN: 1359-5237            Impact factor:   1.444


  10 in total

1.  Clinic, Home, and Kiosk Blood Pressure Measurements for Diagnosing Hypertension: a Randomized Diagnostic Study.

Authors:  Beverly B Green; Melissa L Anderson; Andrea J Cook; Kelly Ehrlich; Yoshio N Hall; Clarissa Hsu; Dwayne Joseph; Predrag Klasnja; Karen L Margolis; Jennifer B McClure; Sean A Munson; Mathew J Thompson
Journal:  J Gen Intern Med       Date:  2022-03-03       Impact factor: 6.473

2.  Implementation of a New Kiosk Technology for Blood Pressure Management in a Family Medicine Clinic: from the WWAMI Region Practice and Research Network.

Authors:  Chia-Fang Chung; Sean A Munson; Matthew J Thompson; Laura-Mae Baldwin; Jeffrey Kaplan; Randall Cline; Beverly B Green
Journal:  J Am Board Fam Med       Date:  2016 Sep-Oct       Impact factor: 2.657

3.  Mean mid-arm circumference and blood pressure cuff sizes for US children, adolescents and adults: National Health and Nutrition Examination Survey, 2011-2016.

Authors:  Yechiam Ostchega; Jeffery P Hughes; Tatiana Nwankwo; Guangyu Zhang
Journal:  Blood Press Monit       Date:  2018-12       Impact factor: 1.444

4.  Blood pressure checks and diagnosing hypertension (BP-CHECK): Design and methods of a randomized controlled diagnostic study comparing clinic, home, kiosk, and 24-hour ambulatory BP monitoring.

Authors:  Beverly B Green; Melissa L Anderson; Jerry Campbell; Andrea J Cook; Kelly Ehrlich; Sarah Evers; Yoshio N Hall; Clarissa Hsu; Dwayne Joseph; Predrag Klasnja; Karen L Margolis; Jennifer B McClure; Sean A Munson; Mathew J Thompson
Journal:  Contemp Clin Trials       Date:  2019-01-08       Impact factor: 2.226

Review 5.  Sources of inaccuracy in the measurement of adult patients' resting blood pressure in clinical settings: a systematic review.

Authors:  Noa Kallioinen; Andrew Hill; Mark S Horswill; Helen E Ward; Marcus O Watson
Journal:  J Hypertens       Date:  2017-03       Impact factor: 4.844

6.  Descriptive analysis of pharmacy services provided after community pharmacy screening.

Authors:  Karla Lancaster; Lehana Thabane; Jean-Eric Tarride; Gina Agarwal; Jeff S Healey; Roopinder Sandhu; Lisa Dolovich
Journal:  Int J Clin Pharm       Date:  2018-11-26

7.  Validation of an Automated and Adjustable Blood Pressure System for Use with a Public Health Station.

Authors:  Timothy R Macaulay; David N Erceg; Bjorn McMillan; Joel E Ramirez; Jesus F Dominguez; Alberto F Vallejo; E Todd Schroeder
Journal:  Vasc Health Risk Manag       Date:  2020-04-09

8.  Public-use blood pressure machines in pharmacies for identification of undetected hypertension in the community.

Authors:  Sherilyn K D Houle; Ross T Tsuyuki
Journal:  J Clin Hypertens (Greenwich)       Date:  2013-01-09       Impact factor: 3.738

9.  Validation of four automatic devices for self-measurement of blood pressure according to the International Protocol of the European Society of Hypertension.

Authors:  Jalil Belghazi; Ramzi N El Feghali; Thérèse Moussalem; Maya Rejdych; Roland G Asmar
Journal:  Vasc Health Risk Manag       Date:  2007

10.  Hypertensive patients' use of blood pressure monitors stationed in pharmacies and other locations: a cross-sectional mail survey.

Authors:  Anthony J Viera; Lauren W Cohen; C Madeline Mitchell; Philip D Sloane
Journal:  BMC Health Serv Res       Date:  2008-10-22       Impact factor: 2.655

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.