Norman Rüb1, Oliver Schweitzer, Christian Mewis, Klaus Kettering, Volker Kuehlkamp.
Abstract
Transvenous unipolar active can defibrillation systems have proven to be effective in treating ventricular tachyarrhythmias. However, a further reduction of ventricular defibrillation thresholds (V-DFT) would increase the longevity, reduce the size of pulse generators, and help to avoid additional leads in patients with inacceptable high V-DFTs. In a finite difference computer model, the extension of the right ventricular (RV) defibrillation coil into the low right atrium led to a 40% reduction of unipolar V-DFT. To evaluate this finding, we conducted a prospective, randomized study in 11 patients receiving an ICD. Extension of the RV electrode was simulated by adding a second coil placed in the low right atrium with the same polarity. Using a binary search protocol, V-DFT was determined with and without the additional electrode in each patient. Total shock impedance was significantly lower in the two coil (low RA) configuration, compared to the single coil (RV) configuration. Corresponding values were 49.9 +/- 6.7 Ohm and 61.1 +/- 9.3 Ohm, respectively (P < 0.01, paired t-test). However, there was no reduction, but even a nonsignificant increase in V-DFTs. Mean V-DFT in the RV configuration was 12.0 +/- 5.6 J and 16.3 +/- 7.8 J in the low RA configuration (P = 0.09, paired t-test). Despite a reduction in total impedance, the addition of a defibrillation coil in the low right atrium does not reduce ventricular defibrillation thresholds.Entities:
Mesh:
Year: 2004 PMID: 15009861 DOI: 10.1111/j.1540-8159.2004.00439.x
Source DB: PubMed Journal: Pacing Clin Electrophysiol ISSN: 0147-8389 Impact factor: 1.976