Preload loss in gold prosthesis-retaining screws as a function of time.

Andrew Cantwell1, John A Hobkirk.   

Abstract

PURPOSE: Screwed joints are widely used in implant dentistry, and their failure is a significant problem that may be related to loss of preload. Preload is the compressive force generated across a joint when a screw is tightened and is responsible for keeping a joint closed. For a given torque, preload is limited by the frictional resistance of the contacting screw threads, flange, and opposing joint surfaces. This study tested the hypothesis that following correct placement, prosthetic gold screws lose preload over time.
MATERIALS AND METHODS: The study used standard Nobel Biocare components. Strain gauges mounted on a standard abutment formed a transducer to measure preload. Five sets of new prosthetic gold screws, gold cylinders, and standard abutment screws were assembled in turn on the top of an implant body, using the transducer abutment. The gold screws were tightened with a Nobel Biocare Torque Controller set to 10 Ncm, when it had an output of 12.06 +/- 0.8 Ncm with 95% repeatability. Preload was monitored for 15 hours; then the screws were removed and examined under a scanning electron microscope.
RESULTS: Preload ranged from 157.5 to 488.9 N (mean 319.6 N), with a mean reduction over 15 hours of 24.9%, with 40.2% of this occurring within 10 seconds of tightening. DISCUSSION: Torsional relaxation of the screw shaft, embedment relaxation, and localized plastic deformation of the gold alloy and opposing titanium threads were the most likely explanation for this phenomenon.
CONCLUSIONS: New prosthetic gold screws suffer significant loss of preload following placement.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14982365

Source DB:  PubMed          Journal:  Int J Oral Maxillofac Implants        ISSN: 0882-2786            Impact factor:   2.804


  8 in total

1.  Effect of cement washout on loosening of abutment screws and vice versa in screw- and cement- retained implant-supported dental prosthesis.

Authors:  Seok-Gyu Kim; Chae-Heon Chung; Mee-Kyoung Son
Journal:  J Adv Prosthodont       Date:  2015-06-23       Impact factor: 1.904

2.  Loosening torque of Universal Abutment screws after cyclic loading: influence of tightening technique and screw coating.

Authors:  Atais Bacchi; Alexandre Regalin; Claudia Lopes Brilhante Bhering; Rodrigo Alessandretti; Aloisio Oro Spazzin
Journal:  J Adv Prosthodont       Date:  2015-10-28       Impact factor: 1.904

3.  Abrasion of abutment screw coated with TiN.

Authors:  Seok-Won Jung; Mee-Kyoung Son; Chae-Heon Chung; Hee-Jung Kim
Journal:  J Adv Prosthodont       Date:  2009-07-31       Impact factor: 1.904

4.  Preload and torque removal evaluation of three different abutment screws for single standing implant restorations.

Authors:  Rafael Augusto Stüker; Eduardo Rolim Teixeira; João Carlos Pinheiro Beck; Nilza Pereira da Costa
Journal:  J Appl Oral Sci       Date:  2008 Jan-Feb       Impact factor: 2.698

5.  Influence of the implant-abutment connection design and diameter on the screw joint stability.

Authors:  Hyon-Mo Shin; Jung-Bo Huh; Mi-Jeong Yun; Young-Chan Jeon; Brian Myung Chang; Chang-Mo Jeong
Journal:  J Adv Prosthodont       Date:  2014-04-22       Impact factor: 1.904

6.  The effect of repeated torque tightening on total lengths of implant abutments in different internal implant‒abutment connections.

Authors:  Fariba Saleh Saber; Nader Abolfazli; Soheil Jannatii Ataei; Mahsa Taghizade Motlagh; Vahede Gharekhani
Journal:  J Dent Res Dent Clin Dent Prospects       Date:  2017-06-21

Review 7.  Implant Bio-mechanics for Successful Implant Therapy: A Systematic Review.

Authors:  Khaled Mosfer Alzahrani
Journal:  J Int Soc Prev Community Dent       Date:  2020-11-24

8.  Comparison of the Effect of Four Different Abutment Screw Torques on Screw Loosening in Single Implant-Supported Prosthesis after the Application of Mechanical Loading.

Authors:  Behnaz Ebadian; Amirhossein Fathi; Saba Khodadad
Journal:  Int J Dent       Date:  2021-07-19
  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.