Estimation of the size of air emboli detectable by electrical impedance measurement.

S Nebuya1, M Noshiro, B H Brown, R H Smallwood, P Milnes.   

Abstract

Non-invasive detection of air emboli in blood is investigated in vitro using a tetrapolar electrical impedance measurement. A cubic tank with a linear array of four electrodes, spaced approximately 1 cm apart down one side, was filled with 0.2 Sm(-1) saline. Bubbles were generated by carbon dioxide gas. Electrical transfer impedance was measured every 8.2 ms at 1.25 MHz. The movement of bubbles was recorded by a video camera, and their sizes and depths from the middle of the array were measured using captured video images. Changes in transfer impedance caused by passage of bubbles were clearly observed and almost identical with those calculated theoretically. Using lead field theory and experimental results, the fundamental limit on the detectable size of bubbles was estimated at the carotid artery, the great saphenous vein and the cephalic vein. The theoretical results showed that a 0.5 mm diameter bubble is detectable at a depth of 5.3 mm, similar to the depth of the great saphenous and the cephalic veins, and a 2.3 mm diameter bubble is detectable at a depth of 21 mm, similar to the depth of the common carotid artery.

Entities:  

Mesh:

Year:  2004        PMID: 14977236     DOI: 10.1007/bf02351024

Source DB:  PubMed          Journal:  Med Biol Eng Comput        ISSN: 0140-0118            Impact factor:   2.602


  8 in total

1.  Decompression limits for compressed air determined by ultrasonically detected blood bubbles.

Authors:  M P Spencer
Journal:  J Appl Physiol       Date:  1976-02       Impact factor: 3.531

2.  An application of electrocardiographic lead theory to impedance plethysmography.

Authors:  D B Geselowitz
Journal:  IEEE Trans Biomed Eng       Date:  1971-01       Impact factor: 4.538

3.  Experiments on decompression bubbles in the circulation using ultrasonic and electromagnetic flowmeters.

Authors:  M P Spencer; S D Campbell; J L Sealey; F C Henry; J Lindbergh
Journal:  J Occup Med       Date:  1969-05

4.  Detection of middle cerebral artery emboli during carotid endarterectomy using transcranial Doppler ultrasonography.

Authors:  M P Spencer; G I Thomas; S C Nicholls; L R Sauvage
Journal:  Stroke       Date:  1990-03       Impact factor: 7.914

5.  Qualitative and quantitative methods for investigating gas emboli in blood.

Authors:  P D Richardson
Journal:  Med Instrum       Date:  1985 Mar-Apr

6.  A comparison of ventilatory and cardiac related changes in EIT images of normal human lungs and of lungs with pulmonary emboli.

Authors:  A D Leathard; B H Brown; J Campbell; F Zhang; A H Morice; D Tayler
Journal:  Physiol Meas       Date:  1994-05       Impact factor: 2.833

7.  Differentiation between different pathological cerebral embolic materials using transcranial Doppler in an in vitro model.

Authors:  H S Markus; M M Brown
Journal:  Stroke       Date:  1993-01       Impact factor: 7.914

8.  Three-dimensional electrical impedance tomography.

Authors:  P Metherall; D C Barber; R H Smallwood; B H Brown
Journal:  Nature       Date:  1996-04-11       Impact factor: 49.962

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.