Total viable counts, ATP, and endotoxin levels as potential markers of microbial contamination of dental unit water systems.

M R Fulford1, J T Walker, M V Martin, P D Marsh.   

Abstract

OBJECTIVES: To determine if either ATP or endotoxin concentrations in water supplied by dental unit water systems (DUWS) correlated with total viable counts (TVC), and therefore could be used as a rapid, chairside measure of levels of microbial contamination.
DESIGN: A prospective trial.
METHOD: Fifty-seven water samples were taken from the 'triple spray', air rotor and source water supplies from 25 dental units in eight practices. The samples were assayed for endotoxin concentration, total ATP and TVC. A pilot study was performed to assess the relationship between TVC and total cell counts, as determined by flow cytometry.
RESULTS: ATP concentrations ranged from 22 to 958 relative light units (RLU) and free endotoxin ranged from 25 to 600 EU ml(-1). TVC varied from not detected to 2.16 x 10(4) CFU ml(-1). The ATP method proved to be a simple and rapid method that could be used at the chairside. However, there was no correlation between ATP or endotoxin concentrations and TVC in DUWS. TVC generally underestimated the total cell count by 50 to 500 fold.
CONCLUSION: Half of the water samples from DUWS exceeded recommended levels of TVC. However, ATP and endotoxin concentrations in DUWS water samples did not correlate with these TVC data and therefore could not be recommended as an alternative assay to TVC for measuring bacterial contamination or for monitoring water treatment efficacy.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14963441     DOI: 10.1038/sj.bdj.4810943

Source DB:  PubMed          Journal:  Br Dent J        ISSN: 0007-0610            Impact factor:   1.626


  4 in total

1.  Workplace determinants of endotoxin exposure in dental healthcare facilities in South Africa.

Authors:  Tanusha S Singh; Braimoh Bello; Onnicah D Mabe; Kevin Renton; Mohamed F Jeebhay
Journal:  Ann Occup Hyg       Date:  2009-12-31

2.  Comparison of the microbial load of incoming and distal outlet waters from dental unit water systems in Istanbul.

Authors:  Irfan Türetgen; Duygu Göksay; Aysin Cotuk
Journal:  Environ Monit Assess       Date:  2008-10-09       Impact factor: 2.513

Review 3.  Monitoring dental-unit-water-line output water by current in-office test kits.

Authors:  Sham Lal; Sim K Singhrao; Matt Bricknell; Mark Pearce; L H Glyn Morton; Waqar Ahmed; St John Crean
Journal:  Curr Microbiol       Date:  2014-03-25       Impact factor: 2.188

4.  Disinfection of dental unit water line using aloe vera: in vitro study.

Authors:  Sonia Pareek; Anup Nagaraj; Prateek Sharma; Mansi Atri; Satinder Walia; Shravani Naidu; Asif Yousuf
Journal:  Int J Dent       Date:  2013-09-08
  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.