Literature DB >> 14871179

The accuracy of electronic working length determination.

D Hoer1, T Attin.   

Abstract

AIM: To determine in vivo the accuracy of two impedance quotient apex locators under clinical conditions.
METHODOLOGY: Electronic working length determination was carried out before extraction in 79 human teeth with 93 root canals. In 51 root canals, the determination was performed using the apex locator Justy II(R) (Hager & Werken, Duisburg, Germany); in 42 canals, the apex locator Endy 5000(R) (Loser, Leverkusen, Germany) was used. A root canal instrument was fixed at working length with composite material prior to extraction followed by the exposure of a radiograph. After histological preparation of the apical region, the teeth were examined under a light microscope. The distance of the file tips to the target intervals 'minor foramen - major foramen' and 'apical canal constriction' was determined. These values were compared with the calculated working lengths, determined by radiographic assessment. The data were statistically analysed by a paired t-test.
RESULTS: For both apex locators and both target intervals, no significant differences between the electronic and radiographical assessments were recorded. The probability of determining the area between minor and major foramen was 82.4% for Justy II and 81% for Endy 5000. However, accurate determination of the apical constriction was only successful in 51% (Justy II) and 64.3% (Endy 5000) of canals. Variation of the inaccurate measurements was higher for Endy 5000 than for Justy II.
CONCLUSIONS: Under clinical conditions, it is possible to determine the region between the minor and major apical foramen with electronic length measuring devices (ELD). However, use of these devices does not result in precise determination of the apical constriction.

Entities:  

Mesh:

Year:  2004        PMID: 14871179     DOI: 10.1111/j.0143-2885.2004.00764.x

Source DB:  PubMed          Journal:  Int Endod J        ISSN: 0143-2885            Impact factor:   5.264


  21 in total

1.  Clinical evaluation of Root ZX II electronic apex locator in primary teeth.

Authors:  S Saritha; K S Uloopi; C Vinay; R Chandra Sekhar; V V Rao
Journal:  Eur Arch Paediatr Dent       Date:  2012-02

2.  The use of high-resolution digital imaging technology for small diameter K-file length determination in endodontics.

Authors:  Bart Vandenberghe; Marius Bud; Asti Sutanto; Reinhilde Jacobs
Journal:  Clin Oral Investig       Date:  2009-05-19       Impact factor: 3.573

3.  The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.

Authors:  Carlos G Adorno; Sandra M Solaeche; Ileana E Ferreira; Andrea Pedrozo; Patricia M Escobar; Vicente R Fretes
Journal:  Clin Oral Investig       Date:  2021-02-13       Impact factor: 3.573

4.  Precision of multi-frequency electronic apex locators.

Authors:  Roy George
Journal:  Evid Based Dent       Date:  2016-09

5.  In vitro evaluation of the accuracy of two electronic apex locators.

Authors:  Viresh Chopra; Shibani Grover; S Datta Prasad
Journal:  J Conserv Dent       Date:  2008-04

6.  Efficiency of 2 electronic apex locators on working length determination: A clinical study.

Authors:  Sibel Koçak; Mustafa Murat Koçak; Baran Can Sağlam
Journal:  J Conserv Dent       Date:  2013-05

7.  In vivo assessment of the impedance ratio method used in electronic foramen locators.

Authors:  Marcos V H Rambo; Humberto R Gamba; Gustavo B Borba; Joaquim M Maia; Carlos A S Ramos
Journal:  Biomed Eng Online       Date:  2010-09-06       Impact factor: 2.819

8.  Preclinical endodontic teaching. A survey of Saudi dental schools.

Authors:  Udaya Narayanaraopeta; Emad Alshwaimi
Journal:  Saudi Med J       Date:  2015-01       Impact factor: 1.484

9.  Survey of attitudes, materials and methods preferred in root canal therapy by general dental practice in Turkey: Part 1.

Authors:  Gul Celik Unal; Bulem Ureyen Kaya; Ali Gurhan Tac; Ayse Diljin Kececi
Journal:  Eur J Dent       Date:  2012-10

10.  Comparison between two methods of working length determination and its effect on radiographic extent of root canal filling: a clinical study [ISRCTN71486641].

Authors:  L Smadi
Journal:  BMC Oral Health       Date:  2006-02-11       Impact factor: 2.757

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.