| Literature DB >> 1485648 |
E A Pastorello1, C Incorvaia, V Pravettoni, A Marelli, L Farioli, M Ghezzi.
Abstract
We investigated the diagnostic value of a new in vitro test, Pharmacia CAP System (Pharmacia Diagnostics AB, Uppsala, Sweden), for the quantitative measurement of allergen-specific IgE antibodies by comparison with RAST in 2 groups of patients, 71 atopic and 48 non-atopic. In the last 20 years RAST has supplied a good diagnostic tool, but this test presents some problems, the main one being sensitivity. The new test has a solid phase able to bind even very small amounts of specific IgE and an anti-IgE tracer with very low cross-reactivity with other immunoglobulins, thus presenting more favourable conditions. From the analysis of our results, Pharmacia CAP System gave higher sensitivity (94% compared to 88% of RAST) with no loss of specificity (96% for both tests). The reliability of these results is ensured by the proper selection of patients who were all suffering from pollinosis and were clinically diagnosed as certainly hypersensitive to a single pollen. A positive trend was found between severity of asthma and levels of specific IgE for timothy. Pharmacia CAP System appears to identify a larger number of atopic patients than RAST.Entities:
Mesh:
Substances:
Year: 1992 PMID: 1485648 DOI: 10.1111/j.1398-9995.1992.tb00665.x
Source DB: PubMed Journal: Allergy ISSN: 0105-4538 Impact factor: 13.146