Literature DB >> 14794

Evaluation of a new automatic calcium ion analyzer.

S Madsen, K Olgaard.   

Abstract

We evaluated a new automatic analyzer for ionized calcium (Ca2+), the Orion Model SS-20, based on a flowthrough ion-exchange electrode. Ca2+ was measured in heparinized whole blood and serum. Within-day variation was 1.2%, day-to-day variation1.6%, and analytical recovery 92.4%. Over the phsiological range interference by K+ and Mg2+ was negligible; major changes in ionic strength, induced by changes in N+ concentration, made correction for a sodium error necessary. Within the physiological range, Ca2+ was inversely correlated to variations in pH. Therefore, to compare Ca2+ values, correction to an apparent pH of 7.40 should be made. The calcium binding effect of heparin was negligible when minimal (4.4 int. units/ml) concentrations of heparin were used. Storage of serum at 4 degrees C for one week resulted in a 4 degrees decrease in apparent serum Ca2+, primarily owing to an increase in pH during storage. In normal material, mean values for blood-and serum-Ca2+(1.10 and 1.07 mmol/liter, respectively) were close to results obtained by previous systems. Errors caused by disturbances in the fluid flow and non-function of half the electrodes we received were the major inconveniences of the analyzer. We conclude that this new analyzer gives decisive advantages in measurement of Ca2+, making this importnant analysis possible as a routine laboratory test for the first time.

Entities:  

Mesh:

Substances:

Year:  1977        PMID: 14794

Source DB:  PubMed          Journal:  Clin Chem        ISSN: 0009-9147            Impact factor:   8.327


  5 in total

1.  Serum ionised calcium concentration: measurement versus calculation.

Authors:  S C Conceicao; D Weightman; P A Smith; J Luno; M K Ward; D N Kerr
Journal:  Br Med J       Date:  1978-04-29

2.  Simultaneous measurements of calcium oxalate crystal nucleation and aggregation: impact of various modifiers.

Authors:  B Hess; U Meinhardt; L Zipperle; R Giovanoli; P Jaeger
Journal:  Urol Res       Date:  1995

3.  Dose calcium balanced heparin affect blood gas and electrolyte analysis?

Authors:  R Ochiai; M Nagano
Journal:  J Anesth       Date:  1990-01       Impact factor: 2.078

4.  Ionised calcium in pathological human bile.

Authors:  D J Sutor; L I Wilkie; M J Jackson
Journal:  J Clin Pathol       Date:  1980-01       Impact factor: 3.411

5.  Therapeutic effect of exercise on hypertension.

Authors:  P E Paulev; R Jordal; O Kristensen; J Ladefoged
Journal:  Eur J Appl Physiol Occup Physiol       Date:  1984
  5 in total