Ang Li1, Quan Zhang, Joseph P Culver, Eric L Miller, David A Boas.
Abstract
We present an algorithm to reconstruct chromosphere concentration images directly rather than following the traditional two-step process of reconstructing wavelength-dependent absorption coefficient images and then calculating chromosphere concentration images. This procedure imposes prior spectral information into the image reconstruction that results in a dramatic improvement in the image contrast-to-noise ratio of better than 100%. We demonstrate this improvement with simulations and a dynamic blood phantom experiment.Mesh:
Substances:
Year: 2004 PMID: 14759043 DOI: 10.1364/ol.29.000256
Source DB: PubMed Journal: Opt Lett ISSN: 0146-9592 Impact factor: 3.776