Using Monte Carlo simulations to commission photon beam output factors--a feasibility study.

George X Ding1.   

Abstract

This study investigates the feasibility of using Monte Carlo methods to assist the commissioning of photon beam output factors from a medical accelerator. The Monte Carlo code, BEAMnrc, was used to model 6 MV and 18 MV photon beams from a Varian linear accelerator. When excellent agreements were obtained between the Monte Carlo simulated and measured dose distributions in a water phantom, the entire geometry including the accelerator head and the water phantom was simulated to calculate the relative output factors. Simulated output factors were compared with measured data, which consist of a typical commission dataset for the output factors. The measurements were done using an ionization chamber in a water phantom at a depth of 10 cm with a source-detector distance of 100 cm. Square fields and rectangular fields with widths and lengths ranging from 4 cm to 40 cm were studied. The result shows a very good agreement (< 1.5%) between the Monte Carlo calculated and the measured relative output factors for a typical commissioning dataset. The Monte Carlo calculated backscatter factors to the beam monitor chamber agree well with measured data in the literature. Monte Carlo simulations have also been shown to be able to accurately predict the collimator exchange effect and its component for rectangular fields. The information obtained is also useful to develop an algorithm for accurate beam modelling. This investigation indicates that Monte Carlo methods can be used to assist commissioning of output factors for photon beams.

Entities:  

Mesh:

Year:  2003        PMID: 14703163     DOI: 10.1088/0031-9155/48/23/005

Source DB:  PubMed          Journal:  Phys Med Biol        ISSN: 0031-9155            Impact factor:   3.609


  10 in total

1.  Monte Carlo calculations for absolute dosimetry to determine machine outputs for proton therapy fields.

Authors:  Harald Paganetti
Journal:  Phys Med Biol       Date:  2006-05-17       Impact factor: 3.609

2.  Monte Carlo simulation for Neptun 10 PC medical linear accelerator and calculations of output factor for electron beam.

Authors:  Mohammad Taghi Bahreyni Toossi; Mehdi Momennezhad; Seyed Mohammad Hashemi
Journal:  Rep Pract Oncol Radiother       Date:  2012-03-06

3.  Total scatter factors of small beams: a multidetector and Monte Carlo study.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon
Journal:  Med Phys       Date:  2008-02       Impact factor: 4.071

4.  Spectral method for the correction of the Cerenkov light effect in plastic scintillation detectors: a comparison study of calibration procedures and validation in Cerenkov light-dominated situations.

Authors:  Mathieu Guillot; Luc Gingras; Louis Archambault; Sam Beddar; Luc Beaulieu
Journal:  Med Phys       Date:  2011-04       Impact factor: 4.071

5.  Development and reproducibility evaluation of a Monte Carlo-based standard LINAC model for quality assurance of multi-institutional clinical trials.

Authors:  Muhammad Nauman Usmani; Hideki Takegawa; Masaaki Takashina; Hodaka Numasaki; Masaki Suga; Yusuke Anetai; Keita Kurosu; Masahiko Koizumi; Teruki Teshima
Journal:  J Radiat Res       Date:  2014-06-23       Impact factor: 2.724

6.  Clinical commissioning and use of the Novalis Tx linear accelerator for SRS and SBRT.

Authors:  Jinkoo Kim; Ning Wen; Jian-Yue Jin; Nicole Walls; Sangroh Kim; Haisen Li; Lei Ren; Yimei Huang; Anthony Doemer; Kathleen Faber; Tina Kunkel; Ahssan Balawi; Kimberly Garbarino; Kenneth Levin; Samir Patel; Munther Ajlouni; Brett Miller; Teamor Nurushev; Calvin Huntzinger; Raymond Schulz; Indrin J Chetty; Benjamin Movsas; Samuel Ryu
Journal:  J Appl Clin Med Phys       Date:  2012-05-10       Impact factor: 2.102

7.  Implementation of a double Gaussian source model for the BEAMnrc Monte Carlo code and its influence on small fields dose distributions.

Authors:  Edgardo Doerner; Paola Caprile
Journal:  J Appl Clin Med Phys       Date:  2016-09-08       Impact factor: 2.102

8.  A multi-centre analytical study of small field output factor calculations in radiotherapy.

Authors:  Krzysztof Chełmiński; Wojciech Bulski
Journal:  Phys Imaging Radiat Oncol       Date:  2018-04-24

9.  Application of a Monte Carlo-based method for total scatter factors of small beams to new solid state micro-detectors.

Authors:  Paolo Francescon; Stefania Cora; Carlo Cavedon; Paolo Scalchi
Journal:  J Appl Clin Med Phys       Date:  2009-01-27       Impact factor: 2.102

10.  Monte Carlo modeling of HD120 multileaf collimator on Varian TrueBeam linear accelerator for verification of 6X and 6X FFF VMAT SABR treatment plans.

Authors:  Alanah M Bergman; Ermias Gete; Cheryl Duzenli; Tony Teke
Journal:  J Appl Clin Med Phys       Date:  2014-05-08       Impact factor: 2.102

  10 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.