BACKGROUND/ PURPOSE: The authors created a new artificial anal sphincter using a shape memory alloy (AS-SMA) to treat fecal incontinence and evaluated its validity. METHODS: AS-SMA consists of 2 Ti-Ni plates to sandwich the intestine and generates a pressure of 55 mm Hg at its resting position. With the electric power supply, the 2 metals bend to form an almondlike shape making a maximum gap of 33 mm between each other at the temperature of 55 degrees C. The device was attached to the colostomy in a piglet and was operated several times a day for 1 month. Fecal continence, bowel movements, and general condition of the piglet were recorded. After the experiment, tissue damage around the device was investigated. RESULTS: Fecal continence was obtained while the device was in the resting position. When it was operated, bowel movements were observed. The bowel movements to operations ratio was 82:105 (78%). During the experiment, the animal had neither abdominal distension nor vomiting. At the autopsy, there was mild inflammation and shallow burns around the device. No compression injury was detected. CONCLUSIONS: AS-SMA achieved fecal continence of the colostomy. With reduction of the associated side effects, it would be a potential substitute for an impaired anal sphincter.
BACKGROUND/ PURPOSE: The authors created a new artificial anal sphincter using a shape memory alloy (AS-SMA) to treat fecal incontinence and evaluated its validity. METHODS: AS-SMA consists of 2 Ti-Ni plates to sandwich the intestine and generates a pressure of 55 mm Hg at its resting position. With the electric power supply, the 2 metals bend to form an almondlike shape making a maximum gap of 33 mm between each other at the temperature of 55 degrees C. The device was attached to the colostomy in a piglet and was operated several times a day for 1 month. Fecal continence, bowel movements, and general condition of the piglet were recorded. After the experiment, tissue damage around the device was investigated. RESULTS: Fecal continence was obtained while the device was in the resting position. When it was operated, bowel movements were observed. The bowel movements to operations ratio was 82:105 (78%). During the experiment, the animal had neither abdominal distension nor vomiting. At the autopsy, there was mild inflammation and shallow burns around the device. No compression injury was detected. CONCLUSIONS: AS-SMA achieved fecal continence of the colostomy. With reduction of the associated side effects, it would be a potential substitute for an impaired anal sphincter.
Authors: Elisa Fattorini; Tobia Brusa; Christian Gingert; Simone E Hieber; Vanessa Leung; Bekim Osmani; Marco D Dominietto; Philippe Büchler; Franc Hetzer; Bert Müller Journal: Ann Biomed Eng Date: 2016-02-29 Impact factor: 3.934