The response of a spherical tissue-equivalent proportional counter to different ions having similar linear energy transfer.

Stephen B Guetersloh1, Thomas B Borak, Phillip J Taddei, Cary Zeitlin, Lawrence Heilbronn, Jack Miller, Takeshi Murakami, Yoshiyuki Iwata.   

Abstract

The response of a tissue-equivalent proportional counter (TEPC) to different ions having a similar linear energy transfer (LET) has been studied. Three ions, 14N, 20Ne and 28Si, were investigated using the HIMAC accelerator at the National Institute of Radiological Sciences at Chiba, Japan. The calculated linear energy transfer (LET( infinity )) of all ions was 44 +/- 2 keV/microm at the sensitive volume of the TEPC. A particle spectrometer was used to record the charge and position of each incident beam particle. This enabled reconstruction of the location of the track as it passed though the TEPC and ensured that the particle survived without fragmentation. The spectrum of energy deposition events in the TEPC could be evaluated as a function of trajectory through the TEPC. The data indicated that there are many events from particles that did not pass through the sensitive volume. The fraction of these events increased as the energy of the particle increased due to changes in the maximum energy of the delta rays. Even though the LET of the incident particles was nearly identical, the frequency-averaged lineal energy, y(F), as well as the dose-averaged lineal energy, y(D), varied with the velocity of the incident particle. However, both values were within 15% of LET in all cases.

Entities:  

Mesh:

Substances:

Year:  2004        PMID: 14680395     DOI: 10.1667/rr3101

Source DB:  PubMed          Journal:  Radiat Res        ISSN: 0033-7587            Impact factor:   2.841


  4 in total

1.  The response of a spherical tissue-equivalent proportional counter to different heavy ions having similar velocities.

Authors:  Phillip J Taddei; Thomas B Borak; Stephen B Guetersloh; Brad B Gersey; Cary Zeitlin; Lawrence Heilbronn; Jack Miller; Takeshi Murakami; Yoshiyuki Iwata
Journal:  Radiat Meas       Date:  2006-10       Impact factor: 1.898

2.  Measurement of the stochastic radial dose distribution for a 30-MeV proton beam using a wall-less tissue-equivalent proportional counter.

Authors:  S Tsuda; T Sato; T Ogawa
Journal:  Radiat Prot Dosimetry       Date:  2015-05-08       Impact factor: 0.972

3.  A comparison of the measured responses of a tissue-equivalent proportional counter to high energy heavy (HZE) particles and those simulated using the Geant4 Monte Carlo code.

Authors:  Phillip J Taddei; Zhongxiang Zhao; Thomas B Borak
Journal:  Radiat Meas       Date:  2008-10       Impact factor: 1.898

4.  Compact Tissue-equivalent Proportional Counter for Deep Space Human Missions.

Authors:  T Straume; L A Braby; T B Borak; T Lusby; D W Warner; D Perez-Nunez
Journal:  Health Phys       Date:  2015-10       Impact factor: 1.316

  4 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.