Literature DB >> 1467040

Transoesophageal Doppler echocardiographic measurement of cardiac output by the mitral annulus method.

H Shimamoto1, H Kito, K Kawazoe, T Fujita, Y Shimamoto.   

Abstract

OBJECTIVE: To compare cardiac output measured by the transoesophageal Doppler and thermodilution techniques.
DESIGN: Prospective direct comparison of paired measurements by both techniques in each patient.
SETTING: Intensive care unit in a cardiovascular centre. PATIENTS: 65 patients after open heart surgery (mean (SD) age 53 (12) years).
INTERVENTIONS: Cardiac output was measured simultaneously by the transoesophageal Doppler and thermodilution techniques. Cardiac output was measured again after a mechanical intervention or volume loading.
RESULTS: The limits of agreement were -2.53 to +0.83 1.min-1 for cardiac output measured by the Doppler and thermodilution techniques. This suggests that the Doppler method alone would not be suitable for clinical use. The second measurement of cardiac output by thermodilution was compared with cardiac output estimated from the first and second Doppler measurements and the first thermodilution measurement. The limits of agreement (-0.55 to +0.51 1.min-1) were good enough for clinical use.
CONCLUSIONS: After cardiac output had been measured simultaneously by both the Doppler and thermodilution techniques, subsequent transoesophageal Doppler alone gave a clinically useful measurement of cardiac output.

Entities:  

Mesh:

Year:  1992        PMID: 1467040      PMCID: PMC1025199          DOI: 10.1136/hrt.68.11.510

Source DB:  PubMed          Journal:  Br Heart J        ISSN: 0007-0769


  15 in total

1.  LEFT RETROGRADE CARDIOANGIOGRAPHY IN ACQUIRED CARDIAC DISEASE: TECHNIC, INDICATIONS AND INTERPRETATIONS IN 700 CASES.

Authors:  R D SELLERS; M J LEVY; K AMPLATZ; C W LILLEHEI
Journal:  Am J Cardiol       Date:  1964-10       Impact factor: 2.778

2.  Analysis of errors in measurement of cardiac output by simultaneous dye and thermal dilution in cardiothoracic surgical patients.

Authors:  A P Fischer; A M Benis; R A Jurado; E Seely; P Teirstein; R S Litwak
Journal:  Cardiovasc Res       Date:  1978-03       Impact factor: 10.787

3.  Statistical methods for assessing agreement between two methods of clinical measurement.

Authors:  J M Bland; D G Altman
Journal:  Lancet       Date:  1986-02-08       Impact factor: 79.321

4.  Evaluation of an elliptical area technique for calculating mitral blood flow by Doppler echocardiography.

Authors:  S J Goldberg; D F Dickinson; N Wilson
Journal:  Br Heart J       Date:  1985-07

5.  Size and motion of the mitral valve annulus in man. I. A two-dimensional echocardiographic method and findings in normal subjects.

Authors:  J A Ormiston; P M Shah; C Tei; M Wong
Journal:  Circulation       Date:  1981-07       Impact factor: 29.690

6.  Color Doppler evaluation of valvular regurgitation in normal subjects.

Authors:  K Yoshida; J Yoshikawa; M Shakudo; T Akasaka; Y Jyo; S Takao; K Shiratori; K Koizumi; F Okumachi; H Kato
Journal:  Circulation       Date:  1988-10       Impact factor: 29.690

7.  Intraoperative detection of myocardial ischemia in high-risk patients: electrocardiography versus two-dimensional transesophageal echocardiography.

Authors:  J S Smith; M K Cahalan; D J Benefiel; B F Byrd; F W Lurz; W A Shapiro; M F Roizen; A Bouchard; N B Schiller
Journal:  Circulation       Date:  1985-11       Impact factor: 29.690

8.  Noninvasive Doppler determination of cardiac output in man. Clinical validation.

Authors:  L L Huntsman; D K Stewart; S R Barnes; S B Franklin; J S Colocousis; E A Hessel
Journal:  Circulation       Date:  1983-03       Impact factor: 29.690

9.  [Mitral regurgitation: detection and quantitative evaluation by two-dimensional Doppler echocardiography].

Authors:  M Ohwa; H Sakakibara; K Miyatake; M Okamoto; N Kinoshita; E Ueda; T Funahashi; I Nakasone; Y Nimura
Journal:  J Cardiogr       Date:  1985-09

10.  A pulsed Doppler echocardiographic method for calculating pulmonary and systemic blood flow in atrial level shunts: validation studies in animals and initial human experience.

Authors:  L M Valdes-Cruz; S Horowitz; E Mesel; D J Sahn; D C Fisher; D Larson
Journal:  Circulation       Date:  1984-01       Impact factor: 29.690

View more
  3 in total

Review 1.  Systematic review of cardiac output measurements by echocardiography vs. thermodilution: the techniques are not interchangeable.

Authors:  Mik Wetterslev; Hasse Møller-Sørensen; Rasmus Rothmann Johansen; Anders Perner
Journal:  Intensive Care Med       Date:  2016-03-01       Impact factor: 17.440

Review 2.  Cardiac output monitoring by echocardiography: should we pass on Swan-Ganz catheters?

Authors:  A C Perrino
Journal:  Yale J Biol Med       Date:  1993 Sep-Oct

3.  Cardiac output measurements via echocardiography versus thermodilution: A systematic review and meta-analysis.

Authors:  Yun Zhang; Yan Wang; Jing Shi; Zhiqiang Hua; Jinyu Xu
Journal:  PLoS One       Date:  2019-10-03       Impact factor: 3.240

  3 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.