Comparison of two spatially resolved NIRS oxygenation indices.

Roy E Gagnon1, Andrew J Macnab, Faith A Gagnon, Derek Blackstock, Jacques G LeBlanc.   

Abstract

UNLABELLED: We compared the percentage haemoglobin oxygenation indices from two near infrared spectrophotometers (NIRS) to determine whether the devices reported similar changes in response to induced changes in oxygenation.
METHODS: 24 healthy juvenile swine undergoing cardiac bypass surgery had INVOS 5100 and NIRO-300 sensors applied to the brow. Induced events included circulatory arrest, altered blood flow rate, core cooling, and re-warming.
RESULTS: The average data collection was 4 hours 36 minutes and had an r = 0.82 mean correlation between the INVOS and NIRO. The total resting baseline collection from all trials (8,590 pairs) had a correlation of r = 0.62. The average relationship between the INVOS and NIRO was non-linear: an INVOS regional oxygen saturation index (rSO2) of 0% was equivalent to a NIRO tissue oxygenation index (TOI) of 36.2%; values were equal at 56.8%; and an (rSO2) of 100% was equivalent to a TOI 85.9%. There was good or excellent agreement (r > 0.5) between the (rSO2) and TOI patterns of change during induced events in 96% of trials. The INVOS and NIRO were most closely correlated when an attenuation filter was used to obtain identical emitter/detector separations.
CONCLUSIONS: There was close agreement between the INVOS 5100 and NIRO-300 in response to major physiological change, although absolute values of (rSO2) and TOI were not identical. There was less agreement during baseline measurements or minimal physiologic change.

Entities:  

Mesh:

Substances:

Year:  2002        PMID: 14650633     DOI: 10.1023/a:1026274124837

Source DB:  PubMed          Journal:  J Clin Monit Comput        ISSN: 1387-1307            Impact factor:   2.502


  24 in total

1.  New instrument that uses near-infrared spectroscopy for the monitoring of human muscle oxygenation.

Authors:  T Muellner; A Nikolic; W Schramm; V Vécsei
Journal:  J Trauma       Date:  1999-06

2.  Near infrared spectroscopy for intraoperative monitoring of the spinal cord.

Authors:  Andrew J Macnab; Roy E Gagnon; Faith A Gagnon
Journal:  Spine (Phila Pa 1976)       Date:  2002-01-01       Impact factor: 3.468

3.  Near infrared spectroscopy (NIRS) distinguishes seizure types.

Authors:  D K Sokol; O N Markand; E C Daly; T G Luerssen; M D Malkoff
Journal:  Seizure       Date:  2000-07       Impact factor: 3.184

4.  Effects of propofol on cerebral oxygenation during cardiopulmonary bypass in children.

Authors:  J G LeBlanc; D Blackstock; A J Macnab; F Gagnon; R Gagnon; J Russell; T Ring
Journal:  Can J Anaesth       Date:  2000-11       Impact factor: 5.063

5.  Validation of transcranial near-infrared spectroscopy for evaluation of cerebral blood flow autoregulation.

Authors:  K S Olsen; L B Svendsen; F S Larsen
Journal:  J Neurosurg Anesthesiol       Date:  1996-10       Impact factor: 3.956

Review 6.  Use of near-infrared spectroscopy in the adult.

Authors:  P J Kirkpatrick
Journal:  Philos Trans R Soc Lond B Biol Sci       Date:  1997-06-29       Impact factor: 6.237

7.  A comparison of the INVOS 4100 and the NIRO 300 near-infrared spectrophotometers.

Authors:  Kenji Yoshitani; Masahiko Kawaguchi; Kazuyuki Tatsumi; Katsuyasu Kitaguchi; Hitoshi Furuya
Journal:  Anesth Analg       Date:  2002-03       Impact factor: 5.108

8.  Optimizing cerebral oxygenation during cardiac surgery: a review of experimental and clinical investigations with near infrared spectrophotometry.

Authors:  G Nollert; R A Jonas; B Reichart
Journal:  Thorac Cardiovasc Surg       Date:  2000-08       Impact factor: 1.827

9.  Cerebral oxygen supply and utilization during infant cardiac surgery.

Authors:  A J du Plessis; J Newburger; R A Jonas; P Hickey; H Naruse; M Tsuji; A Walsh; G Walter; D Wypij; J J Volpe
Journal:  Ann Neurol       Date:  1995-04       Impact factor: 10.422

10.  Monitoring blood loss with near infrared spectroscopy.

Authors:  Francesco Torella; Richard Cowley; Maureen S Thorniley; Charles N McCollum
Journal:  Comp Biochem Physiol A Mol Integr Physiol       Date:  2002-05       Impact factor: 2.320

View more
  8 in total

1.  Calibration of a prototype NIRS oximeter against two commercial devices on a blood-lipid phantom.

Authors:  Simon Hyttel-Sorensen; Stefan Kleiser; Martin Wolf; Gorm Greisen
Journal:  Biomed Opt Express       Date:  2013-08-14       Impact factor: 3.732

2.  Volatile isoflurane sedation in cerebrovascular intensive care patients using AnaConDa(®): effects on cerebral oxygenation, circulation, and pressure.

Authors:  Julian Bösel; Jan C Purrucker; Frank Nowak; Julian Renzland; Petra Schiller; Eva Benveniste Pérez; Sven Poli; Benjamin Brunn; Werner Hacke; Thorsten Steiner
Journal:  Intensive Care Med       Date:  2012-10-25       Impact factor: 17.440

3.  Cerebral oximetry during infant cardiac surgery: evaluation and relationship to early postoperative outcome.

Authors:  Barry D Kussman; David Wypij; James A DiNardo; Jane W Newburger; John E Mayer; Pedro J del Nido; Emile A Bacha; Frank Pigula; Ellen McGrath; Peter C Laussen
Journal:  Anesth Analg       Date:  2009-04       Impact factor: 5.108

4.  Tissue oximetry: a comparison of mean values of regional tissue saturation, reproducibility and dynamic range of four NIRS-instruments on the human forearm.

Authors:  Simon Hyttel-Sorensen; Line C Sorensen; Joan Riera; Gorm Greisen
Journal:  Biomed Opt Express       Date:  2011-10-06       Impact factor: 3.732

5.  Maintaining Gait Performance by Cortical Activation during Dual-Task Interference: A Functional Near-Infrared Spectroscopy Study.

Authors:  Chia-Feng Lu; Yan-Ci Liu; Yea-Ru Yang; Yu-Te Wu; Ray-Yau Wang
Journal:  PLoS One       Date:  2015-06-16       Impact factor: 3.240

Review 6.  Near Infrared Spectroscopy for Poor Grade Aneurysmal Subarachnoid Hemorrhage-A Concise Review.

Authors:  Charles L Francoeur; François Lauzier; Patrice Brassard; Alexis F Turgeon
Journal:  Front Neurol       Date:  2022-04-18       Impact factor: 4.086

7.  Comparing near-infrared spectroscopy-measured cerebral oxygen saturation and corresponding venous oxygen saturations in children with congenital heart disease: a systematic review and meta-analysis.

Authors:  Yiqi Ma; Lihong Zhao; Jiafu Wei; Ziwei Wang; Su Lui; Bin Song; Qiyong Gong; Pu Wang; Min Wu
Journal:  Transl Pediatr       Date:  2022-08

8.  Simultaneous measurement of deep tissue blood flow and oxygenation using noncontact diffuse correlation spectroscopy flow-oximeter.

Authors:  Ting Li; Yu Lin; Yu Shang; Lian He; Chong Huang; Margaret Szabunio; Guoqiang Yu
Journal:  Sci Rep       Date:  2013       Impact factor: 4.379

  8 in total

北京卡尤迪生物科技股份有限公司 © 2022-2023.