Literature DB >> 14646699

Fractures of ceramic bearings: history and present status.

Didier Hannouche1, Christophe Nich, Pascal Bizot, Alain Meunier, Rémi Nizard, Laurent Sedel.   

Abstract

During a period of 25 years (1977 to 2001), 13 fractures of an alumina component were recorded retrospectively. During the same period, approximately 5500 alumina components were implanted (3300 with all-alumina bearings and 1200 with alumina-on-polyethylene). These events occurred in seven women and six men, with a mean age of 59 years and a mean weight of 71 kg. There were eight fractures of the femoral head and five fractures of the socket component. Three fractures clearly were related to trauma (two fractures of the socket and one fracture of the head) and two were related to an abnormal design (one 22-mm head and one extra-long neck). Five fractures occurred without any rational explanation, but two fractures that occurred in the early phase could have been caused by a weaker alumina material. Finally three recent fractures of the liner were related to a change in the design of the material. This has been corrected and no additional fractures were observed. Although this dramatic event is of concern, it is infrequent and easy to solve by a limited revision procedure if done emergently. This rare complication should be balanced with the other complications of THR, including mechanical failures and osteolysis, often described with metal-on-polyethylene prostheses.

Entities:  

Mesh:

Year:  2003        PMID: 14646699     DOI: 10.1097/01.blo.0000096806.78689.50

Source DB:  PubMed          Journal:  Clin Orthop Relat Res        ISSN: 0009-921X            Impact factor:   4.176


  32 in total

1.  Fractures of a sandwich ceramic liner at ten year follow-up.

Authors:  Anthony Viste; Julien Chouteau; Romain Desmarchelier; Michel-Henri Fessy
Journal:  Int Orthop       Date:  2011-10-20       Impact factor: 3.075

2.  The 2009 Frank Stinchfield Award: "Hip squeaking": a biomechanical study of ceramic-on-ceramic bearing surfaces.

Authors:  Christophe Chevillotte; Robert T Trousdale; Qingshan Chen; Olivier Guyen; Kai-Nan An
Journal:  Clin Orthop Relat Res       Date:  2009-06-19       Impact factor: 4.176

Review 3.  Biological response to prosthetic debris.

Authors:  Diana Bitar; Javad Parvizi
Journal:  World J Orthop       Date:  2015-03-18

4.  Clinical Faceoff: Instability After THA: The Potential Role of the Bearing Surface.

Authors:  Javad Parvizi; Laurent Sedel; Michael J Dunbar
Journal:  Clin Orthop Relat Res       Date:  2018-04       Impact factor: 4.176

5.  Alumina-on-alumina THA performed in patients younger than 30 years: a 10-year minimum followup study.

Authors:  Hyeong Jo Yoon; Jeong Joon Yoo; Kang Sup Yoon; Kyung-Hoi Koo; Hee Joong Kim
Journal:  Clin Orthop Relat Res       Date:  2012-07-24       Impact factor: 4.176

6.  Ceramic bearings for total hip arthroplasty have high survivorship at 10 years.

Authors:  James A D'Antonio; William N Capello; Marybeth Naughton
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

7.  High survivorship with a titanium-encased alumina ceramic bearing for total hip arthroplasty.

Authors:  James A D'Antonio; William N Capello; Marybeth Naughton
Journal:  Clin Orthop Relat Res       Date:  2014-02       Impact factor: 4.176

8.  Factors affecting modular acetabular ceramic liner insertion: a biomechanical analysis.

Authors:  James P McAuley; Douglas A Dennis; Justin Grostefon; William G Hamilton
Journal:  Clin Orthop Relat Res       Date:  2012-02       Impact factor: 4.176

9.  THA with Delta ceramic on ceramic: results of a multicenter investigational device exemption trial.

Authors:  William G Hamilton; James P McAuley; Douglas A Dennis; Jeffrey A Murphy; Thomas J Blumenfeld; Joel Politi
Journal:  Clin Orthop Relat Res       Date:  2009-09-19       Impact factor: 4.176

Review 10.  [The problem of fractures of ceramic heads. What should be done?].

Authors:  H R Dürr
Journal:  Orthopade       Date:  2009-08       Impact factor: 1.087

View more