In vivo dosimetry using radiochromic films during intraoperative electron beam radiation therapy in early-stage breast cancer.

Mario Ciocca1, Roberto Orecchia, Cristina Garibaldi, Elena Rondi, Alberto Luini, Giovanna Gatti, Mattia Intra, Paolo Veronesi, Roberta Lazzari, Giampiero Tosi, Umberto Veronesi.   

Abstract

BACKGROUND AND
PURPOSE: To check the dose delivered to patients during intraoperative electron beam radiation therapy (IOERT) for early breast cancer and also to define appropriate action levels. PATIENTS AND METHODS: Between December 2000 and June 2001, 54 patients affected by early-stage breast cancer underwent exclusive IOERT to the tumour bed using a Novac7 mobile linac, after quadrantectomy. Electron beams (5, 7, 9 MeV) at high dose per pulse values (0.02-0.09 Gy/pulse) were used. The prescribed single dose was 21 Gy at the depth of 90% isodose (14-22 mm). In 35 cases, in vivo dosimetry was performed. The entrance dose was derived from the surface dose measured with thin and calibrated MD-55-2 radiochromic films, wrapped in sterile envelopes. Films were analysed 24-72 h after the irradiation using a charge-coupled-device imaging system. Field disturbance caused by the film envelope was negligible.
RESULTS: The mean deviation between measured and expected doses was 1.8%, with one SD equal to 4.7%. Deviations larger than 7% were found in 23% of cases, never consecutively, not correlated with beam energy or field size and with no evidence of linac daily output variation or serious malfunctioning or human mistake. The estimated overall uncertainty of dose measurement was about 4%. In vivo dosimetry appeared both reliable and feasible. Two action levels, for unexplained observed deviations larger than 7 and 10%, were preliminary defined.
CONCLUSIONS: Satisfactory agreement between measured and expected doses was found. The implementation of in vivo dosimetry in IOERT is suggested, particularly for patients enrolled in a clinical trial.

Entities:  

Mesh:

Year:  2003        PMID: 14644488     DOI: 10.1016/j.radonc.2003.09.001

Source DB:  PubMed          Journal:  Radiother Oncol        ISSN: 0167-8140            Impact factor:   6.280


  11 in total

1.  In vivo dosimetry in intraoperative electron radiotherapy: microMOSFETs, radiochromic films and a general-purpose linac.

Authors:  Juan López-Tarjuelo; Ana Bouché-Babiloni; Virginia Morillo-Macías; Noelia de Marco-Blancas; Agustín Santos-Serra; Juan David Quirós-Higueras; Carlos Ferrer-Albiach
Journal:  Strahlenther Onkol       Date:  2014-06-26       Impact factor: 3.621

2.  In vivo dosimetry using Gafchromic films during pelvic intraoperative electron radiation therapy (IOERT).

Authors:  Filipa Costa; Sandra Sarmento; Dora Gomes; Helena Magalhães; Rosário Arrais; Graciete Moreira; Maria Fátima Cruz; José Pedro Silva; Lúcio Santos; Olga Sousa
Journal:  Br J Radiol       Date:  2016-05-18       Impact factor: 3.039

3.  Full-dose intraoperative radiotherapy with electrons during breast-conserving surgery: experience with 590 cases.

Authors:  Umberto Veronesi; Roberto Orecchia; Alberto Luini; Viviana Galimberti; Giovanna Gatti; Mattia Intra; Paolo Veronesi; Maria Cristina Leonardi; Mario Ciocca; Roberta Lazzari; Pietro Caldarella; Serife Simsek; Luzemira Santos Silva; Daniele Sances
Journal:  Ann Surg       Date:  2005-07       Impact factor: 12.969

4.  Full-dose intra-operative radiotherapy with electrons (ELIOT) during breast-conserving surgery: experience with 1246 cases.

Authors:  U Veronesi; R Orecchia; A Luini; V Galimberti; G Gatti; M Intra; P Veronesi; M C Leonardi; M Ciocca; R Lazzari; P Caldarella; N Rotmensz; C Sangalli; L S Silva; D Sances
Journal:  Ecancermedicalscience       Date:  2008-02-26

5.  Evaluation of LiF:Mg,Ti (TLD-100) for Intraoperative Electron Radiation Therapy Quality Assurance.

Authors:  Raffaele Liuzzi; Federica Savino; Vittoria D'Avino; Mariagabriella Pugliese; Laura Cella
Journal:  PLoS One       Date:  2015-10-01       Impact factor: 3.240

6.  Intraoperative full-dose of partial breast irradiation with electrons delivered by standard linear accelerators for early breast cancer.

Authors:  Alfredo Carlos S D Barros; Samir A Hanna; Heloísa A Carvalho; Eduardo Martella; Felipe Eduardo M Andrade; José Roberto M Piato; José Luiz B Bevilacqua
Journal:  Int J Breast Cancer       Date:  2014-12-17

7.  Implementation of an intraoperative electron radiotherapy in vivo dosimetry program.

Authors:  Juan López-Tarjuelo; Virginia Morillo-Macías; Ana Bouché-Babiloni; Enrique Boldó-Roda; Rafael Lozoya-Albacar; Carlos Ferrer-Albiach
Journal:  Radiat Oncol       Date:  2016-03-15       Impact factor: 3.481

8.  In vivo dosimetry and shielding disk alignment verification by EBT3 GAFCHROMIC film in breast IOERT treatment.

Authors:  Mara Severgnini; Mario de Denaro; Marina Bortul; Cristiana Vidali; Aulo Beorchia
Journal:  J Appl Clin Med Phys       Date:  2014-01-08       Impact factor: 2.102

Review 9.  Present state and issues in IORT Physics.

Authors:  Frank W Hensley
Journal:  Radiat Oncol       Date:  2017-01-27       Impact factor: 3.481

10.  Comparing the dosimetric characteristics of the electron beam from dedicated intraoperative and conventional radiotherapy accelerators.

Authors:  Hamid Reza Baghani; Seyed Mahmoud Reza Aghamiri; Seyed Rabi Mahdavi; Mohammad Esmail Akbari; Hamid Reza Mirzaei
Journal:  J Appl Clin Med Phys       Date:  2015-03-08       Impact factor: 2.102

View more

北京卡尤迪生物科技股份有限公司 © 2022-2023.